Effects of preanalytical factors on hemoglobin measurement: A comparison of two HemoCue® point-of-care analyzers.


ABSTRACT: In field studies, hemoglobin (Hb) is often measured using a battery-operated, portable HemoCue® hemoglobinometer.We compared the performance of 2 HemoCue® models (Hb-201+ and Hb-301) and investigated effects of preanalytical factors on Hb results by simulating unfavorable field conditions.The Hb-301 produced 2.6% higher results compared to the Hb-201+. Hb had to be measured within 1min of filling the Hb-301 cuvette to avoid artificially elevated concentrations (1.3% per min). The Hb-301 cuvettes withstood elevated temperature (37°C) and humidity (72%) for 3weeks, while the Hb-201+ cuvettes degraded within 10min under those conditions. Both cuvette types withstood elevated temperature for 3weeks. Properly-collected venous and capillary blood produced comparable results. Pooled capillary blood produced comparable results to the second and third but not the fourth drop of blood (3.3% lower). Blood could be stored for ?4days at 10-30°C before Hb-201+ measurement, but only for 1day at 10-23°C before Hb-301 measurement (?1% change in Hb).Higher Hb results obtained with the Hb-301 may influence the interpretation of anemia prevalence in health surveys. While the Hb-301 performed better in high humidity conditions, the Hb-201+ provided more user flexibility regarding delayed Hb reading.

SUBMITTER: Whitehead RD 

PROVIDER: S-EPMC5685159 | biostudies-literature | 2017 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Effects of preanalytical factors on hemoglobin measurement: A comparison of two HemoCue® point-of-care analyzers.

Whitehead Ralph D RD   Zhang Ming M   Sternberg Maya R MR   Schleicher Rosemary L RL   Drammeh Bakary B   Mapango Carine C   Pfeiffer Christine M CM  

Clinical biochemistry 20170412 9


<h4>Background</h4>In field studies, hemoglobin (Hb) is often measured using a battery-operated, portable HemoCue® hemoglobinometer.<h4>Methods</h4>We compared the performance of 2 HemoCue® models (Hb-201+ and Hb-301) and investigated effects of preanalytical factors on Hb results by simulating unfavorable field conditions.<h4>Results</h4>The Hb-301 produced 2.6% higher results compared to the Hb-201+. Hb had to be measured within 1min of filling the Hb-301 cuvette to avoid artificially elevated  ...[more]

Similar Datasets

| S-EPMC3253126 | biostudies-literature
| S-EPMC6082813 | biostudies-other
| S-EPMC7529235 | biostudies-literature
| S-EPMC6709845 | biostudies-literature
| S-EPMC7160596 | biostudies-literature
| S-EPMC7175548 | biostudies-literature
| S-EPMC8859550 | biostudies-literature
| S-EPMC5814368 | biostudies-literature
| S-EPMC5969348 | biostudies-literature
| S-EPMC6817248 | biostudies-literature