Unknown

Dataset Information

0

Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine.


ABSTRACT: BACKGROUND:Worldwide, ?1 million people manage their type 1 diabetes with an insulin pump and a continuous subcutaneous insulin infusion (CSII) catheter. Patients routinely insert a new catheter every 2-3 days due to increasing variability of insulin absorption over time. Catheter insertion and maintenance damage capillaries, lymphatics, cells, and connective tissue leading to an acute inflammatory response. METHODS:We compared an investigational CSII catheter (IC) and a commercial CSII catheter (CC) regarding insulin absorption pharmacokinetics (PK) and tissue inflammation. The two different catheter designs were implanted into the subcutaneous tissue of six swine for 5 days. Insulin boluses were given on days 1, 3, and 5 of wear-time to assess PK. Tissue around catheters was excised and stained to visualize inflammation and morphological changes of adjacent tissue. RESULTS:Insulin absorption was better when infused through a CC with highest Cmax and fastest tmax values on day 5 of catheter wear-time. Both catheter types produced high intra- and intersubject day-to-day insulin absorption variability. The IC caused significantly more tissue disruption and lead to irregular changes in tissue morphology. Both catheter types were surrounded by a layer of inflammatory tissue that varied in composition, thickness, and density over time. A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion. CONCLUSIONS:Insulin absorption variability could be attributed to the layer of inflammatory tissue, which may function as a mechanical barrier to insulin flow into adjacent vascular tissue. The impact of the acute inflammatory tissue response on insulin absorption has to be considered in future catheter designs. A catheter that was manually inserted by pushing a sharp tip through the skin caused more trauma and variability than a 90° Teflon cannula with automated insertion.

SUBMITTER: Hauzenberger JR 

PROVIDER: S-EPMC5689134 | biostudies-literature | 2017 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Detailed Analysis of Insulin Absorption Variability and the Tissue Response to Continuous Subcutaneous Insulin Infusion Catheter Implantation in Swine.

Hauzenberger Jasmin R JR   Hipszer Brian R BR   Loeum Channy C   McCue Peter A PA   DeStefano Mark M   Torjman Marc C MC   Kaner Mahmut T MT   Dinesen Alek R AR   Chervoneva Inna I   Pieber Thomas R TR   Joseph Jeffrey I JI  

Diabetes technology & therapeutics 20171005 11


<h4>Background</h4>Worldwide, ∼1 million people manage their type 1 diabetes with an insulin pump and a continuous subcutaneous insulin infusion (CSII) catheter. Patients routinely insert a new catheter every 2-3 days due to increasing variability of insulin absorption over time. Catheter insertion and maintenance damage capillaries, lymphatics, cells, and connective tissue leading to an acute inflammatory response.<h4>Methods</h4>We compared an investigational CSII catheter (IC) and a commercia  ...[more]

Similar Datasets

| S-EPMC8080918 | biostudies-literature
| S-EPMC6937034 | biostudies-literature
| S-EPMC8279595 | biostudies-literature
| S-EPMC7196863 | biostudies-literature
| S-EPMC6018811 | biostudies-literature
| S-EPMC8120055 | biostudies-literature
| S-EPMC8245368 | biostudies-literature
| S-EPMC2854665 | biostudies-literature
| S-EPMC5248540 | biostudies-literature
| S-EPMC7048888 | biostudies-literature