Song Kwang Hyun KH Snyder Karen Chin KC Kim Jinkoo J Li Haisen H Ning Wen W Rusnac Robert R Jackson Paul P Gordon James J Siddiqui Salim M SM Chetty Indrin J IJ
Journal of applied clinical medical physics 20160708 4
2.5 MV electronic portal imaging, available on Varian TrueBeam machines, was characterized using various phantoms in this study. Its low-contrast detectability, spatial resolution, and contrast-to-noise ratio (CNR) were compared with those of conventional 6 MV and kV planar imaging. Scatter effect in large patient body was simulated by adding solid water slabs along the beam path. The 2.5 MV imaging mode was also evaluated using clinically acquired images from 24 patients for the sites of brain, ...[more]