Episcleral eye plaque dosimetry comparison for the Eye Physics EP917 using Plaque Simulator and Monte Carlo simulation.


ABSTRACT: This work is a comparative study of the dosimetry calculated by Plaque Simulator, a treatment planning system for eye plaque brachytherapy, to the dosimetry calculated using Monte Carlo simulation for an Eye Physics model EP917 eye plaque. Monte Carlo (MC) simulation using MCNPX 2.7 was used to calculate the central axis dose in water for an EP917 eye plaque fully loaded with 17 IsoAid Advantage (125)I seeds. In addition, the dosimetry parameters Λ, gL(r), and F(r,θ) were calculated for the IsoAid Advantage model IAI-125 (125)I seed and benchmarked against published data. Bebig Plaque Simulator (PS) v5.74 was used to calculate the central axis dose based on the AAPM Updated Task Group 43 (TG-43U1) dose formalism. The calculated central axis dose from MC and PS was then compared. When the M