ABSTRACT: Objective
The purpose of this study was to compare the stability and feasibility of four fixation constructs in a posterior column acetabular fracture: one reconstruction plate, one reconstruction plate and lag screw, two reconstruction plates, and a W-shaped acetabular angular plate.Methods
Twenty embalmed cadaveric pelvises with a posterior column acetabular fractures were allocated to one of four groups: 1) a reconstruction plate, 2) a reconstruction plate with a posterior column lag screw, 3) double reconstruction plates, and 4) a W-shaped acetabular angular plate. These constructs were mechanically loaded on a testing machine, and construct stiffness values were measured. Strain gauges were utilized to measure the mechanical behavior in the condition of compressive force.Results
Final stiffness was not different between the two reconstruction plates (445.81±98.30 N/mm) and the W-shaped acetabular angular plate (447.43±98.45 N/mm, p = 0.524), both of which were superior to a single reconstruction plate (248.90±61.95 N/mm) and a combined plate and lag screw (326.41±94.34 N/mm). Following the fixation of the W-shaped acetabular angular plate, the strain distribution was similar to the intact condition around the acetabulum. The parameters of the W-shaped acetabular angular plate that were observed at the superior region of the acetabulum were less than those of a single reconstruction plate (p<0.05), a single reconstruction plate with lag screw (p<0.05), and two reconstruction plates (p<0.05).Conclusions
The novel W-shaped acetabular angular plate fixation technique was able to provide the biomechanically stiffest construct for stabilization of a posterior column acetabular fracture; it also resulted in a partial restoration of joint loading parameters toward the intact state.
SUBMITTER: Su K
PROVIDER: S-EPMC5695793 | biostudies-literature | 2017
REPOSITORIES: biostudies-literature
PloS one 20171120 11
<h4>Objective</h4>The purpose of this study was to compare the stability and feasibility of four fixation constructs in a posterior column acetabular fracture: one reconstruction plate, one reconstruction plate and lag screw, two reconstruction plates, and a W-shaped acetabular angular plate.<h4>Methods</h4>Twenty embalmed cadaveric pelvises with a posterior column acetabular fractures were allocated to one of four groups: 1) a reconstruction plate, 2) a reconstruction plate with a posterior col ...[more]