Ultrapure dialysis water obtained with additional ultrafilter may reduce inflammation in patients on hemodialysis.


ABSTRACT:

Background

Patients on standard dialysis, in particular those on high-flux and high-efficiency dialysis, are exposed to hundreds of liters of dialysis-water per week. The quality of dialysis-water is a factor responsible for inflammation in dialysis patients. Inflammation is a potent trigger of atherosclerosis and a pathogenetic factor in anemia, increasing mortality and morbidity in dialysis patients. Current systems for water treatment do not completely eliminate bacteria and endotoxins. This prospective study tested whether improved dialysis-water purity by an additional ultrafilter can reduce inflammation and ameliorate hemoglobin levels, with a consequent reduction in erythropoietin-stimulating agents (ESA).

Methods

An ultrafilter, composed of two serially positioned devices with polysulfone membranes of 2.0 and 1.0 m2, respectively, was positioned within the fluid pathway before the dialysis machine. Prevalent dialysis patients were assigned either to continue dialysis with conventional dialysis-water (control phase) or to initiate dialysis sessions with improved dialysis-water purity (study phase). After 6 months, patients were crossed over. Total study duration was 1 year. Routine chemistry, bacterial count, endotoxin levels in dialysis-water as well as blood levels of pro- and anti-inflammatory cytokines, human serum amyloid A, C-reactive protein and fraction 5 of complement were measured.

Results

Thirty-two patients completed the study. Mean bacterial count was lower and endotoxin levels were absent in dialysis-water obtained with the ultrafilter. At the end of the study-phase, C-reactive protein and pro-inflammatory cytokines decreased while anti-inflammatory ones increased. Hemoglobin levels were improved with lower ESA doses.

Conclusions

An additional ultrafilter improved dialysis-water purity, reduced levels of inflammation markers, ameliorated hemoglobin concentration with reduced ESA doses. These results remain speculative but they may generate studies to assess whether improved dialysis-water quality with an ultrafilter can reduce inflammation and improve survival of dialysis patients.

SUBMITTER: Di Iorio B 

PROVIDER: S-EPMC5698401 | biostudies-literature | 2017 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Ultrapure dialysis water obtained with additional ultrafilter may reduce inflammation in patients on hemodialysis.

Di Iorio Biagio B   Di Micco Lucia L   Bruzzese Dario D   Nardone Luca L   Russo Luigi L   Formisano Pietro P   D'Esposito Vittoria V   Russo Domenico D  

Journal of nephrology 20170823 6


<h4>Background</h4>Patients on standard dialysis, in particular those on high-flux and high-efficiency dialysis, are exposed to hundreds of liters of dialysis-water per week. The quality of dialysis-water is a factor responsible for inflammation in dialysis patients. Inflammation is a potent trigger of atherosclerosis and a pathogenetic factor in anemia, increasing mortality and morbidity in dialysis patients. Current systems for water treatment do not completely eliminate bacteria and endotoxin  ...[more]

Similar Datasets

| S-EPMC5234772 | biostudies-literature
| S-EPMC7063748 | biostudies-literature
| S-EPMC6206892 | biostudies-other
| S-EPMC4406314 | biostudies-literature
| S-EPMC6661273 | biostudies-literature
| S-EPMC123900 | biostudies-literature
| S-EPMC8669136 | biostudies-literature
| S-EPMC10907222 | biostudies-literature
| S-EPMC103666 | biostudies-literature
| S-EPMC3097400 | biostudies-literature