Dosimetry of gamma chamber blood irradiator using PAGAT gel dosimeter and Monte Carlo simulations.


ABSTRACT: Currently, the use of blood irradiation for inactivating pathogenic microbes in infected blood products and preventing graft-versus-host disease (GVHD) in immune suppressed patients is greater than ever before. In these systems, dose distribution and uniformity are two important concepts that should be checked. In this study, dosimetry of the gamma chamber blood irradiator model Gammacell 3000 Elan was performed by several dosimeter methods including thermoluminescence dosimeters (TLD), PAGAT gel dosimetry, and Monte Carlo simulations using MCNP4C code. The gel dosimeter was put inside a glass phantom and the TL dosimeters were placed on its surface, and the phantom was then irradiated for 5 min and 27 sec. The dose values at each point inside the vials were obtained from the magnetic resonance imaging of the phantom. For Monte Carlo simulations, all components of the irradiator were simulated and the dose values in a fine cubical lattice were calculated using tally F6. This study shows that PAGAT gel dosimetry results are in close agreement with the results of TL dosimetry, Monte Carlo simulations, and the results given by the vendor, and the percentage difference between the different methods is less than 4% at different points inside the phantom. According to the results obtained in this study, PAGAT gel dosimetry is a reliable method for dosimetry of the blood irradiator. The major advantage of this kind of dosimetry is that it is capable of 3D dose calculation.

SUBMITTER: Mohammadyari P 

PROVIDER: S-EPMC5711240 | biostudies-literature | 2014 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Dosimetry of gamma chamber blood irradiator using PAGAT gel dosimeter and Monte Carlo simulations.

Mohammadyari Parvin P   Zehtabian Mehdi M   Sina Sedigheh S   Tavasoli Ali Reza AR   Faghihi Reza R  

Journal of applied clinical medical physics 20140104 1


Currently, the use of blood irradiation for inactivating pathogenic microbes in infected blood products and preventing graft-versus-host disease (GVHD) in immune suppressed patients is greater than ever before. In these systems, dose distribution and uniformity are two important concepts that should be checked. In this study, dosimetry of the gamma chamber blood irradiator model Gammacell 3000 Elan was performed by several dosimeter methods including thermoluminescence dosimeters (TLD), PAGAT ge  ...[more]

Similar Datasets

| S-EPMC10665335 | biostudies-literature
| S-EPMC4587958 | biostudies-literature
| S-EPMC2999362 | biostudies-literature
| S-EPMC6465329 | biostudies-literature
| S-EPMC9291792 | biostudies-literature
| S-EPMC3808443 | biostudies-other
| S-EPMC6432162 | biostudies-literature
| S-EPMC2998598 | biostudies-literature
| S-EPMC4919768 | biostudies-literature
| S-EPMC8655925 | biostudies-literature