Corrosion and adverse tissue reaction after modular unipolar hip hemiarthroplasty.


ABSTRACT: Modern unipolar hip hemiarthroplasty, commonly used for displaced femoral neck fracture, is now modular, with both a variable length cobalt-chromium adapter-sleeve and large femoral head. Patients with these modular components may develop symptomatic trunnion corrosion, with elevated serum metal levels. We report the case of an 82-year-old woman, 5.5 years after a modular unipolar hip hemiarthroplasty, who presented with a 4-month history of hip pain and limp. Evaluation showed elevated serum cobalt and chromium levels and an acetabular cyst. At revision, fluid, tissue, and gross inspection were consistent with trunnion corrosion. The hip was revised with a ceramic head and dual mobility acetabular component, with a good result at 1 year. The designs of commercially available, modern unipolar hip hemiarthroplasty prostheses are reviewed.

SUBMITTER: Moore BF 

PROVIDER: S-EPMC5712032 | biostudies-literature | 2017 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Corrosion and adverse tissue reaction after modular unipolar hip hemiarthroplasty.

Moore Brian F BF   Lachiewicz Paul F PF  

Arthroplasty today 20170221 4


Modern unipolar hip hemiarthroplasty, commonly used for displaced femoral neck fracture, is now modular, with both a variable length cobalt-chromium adapter-sleeve and large femoral head. Patients with these modular components may develop symptomatic trunnion corrosion, with elevated serum metal levels. We report the case of an 82-year-old woman, 5.5 years after a modular unipolar hip hemiarthroplasty, who presented with a 4-month history of hip pain and limp. Evaluation showed elevated serum co  ...[more]

Similar Datasets

| S-EPMC7772450 | biostudies-literature
| S-EPMC7303482 | biostudies-literature
| S-EPMC7239747 | biostudies-literature
| S-EPMC7370985 | biostudies-literature
| S-EPMC2667129 | biostudies-literature