Comparison of TLD calibration methods for 192Ir dosimetry.


ABSTRACT: For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in (192)Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachytherapy audit where seven Australian centers irradiated three sets of TLDs in a water phantom. The results were in agreement within estimated uncertainties when the TLDs were calibrated with the (192)Ir source. Calibrating TLDs in a phantom similar to that used for the audit proved to be the most practical method and provided the greatest confidence in measured dose. When calibrated using 6 MV photons, the TLD results were consistently higher than the (192)Ir-calibrated TLDs, suggesting this method does not fully correct for the response of the TLDs when irradiated in the audit phantom.

SUBMITTER: Haworth A 

PROVIDER: S-EPMC5714053 | biostudies-literature | 2013 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of TLD calibration methods for 192Ir dosimetry.

Haworth Annette A   Butler Duncan J DJ   Wilfert Lisa L   Ebert Martin A MA   Todd Stephen P SP   Hayton Anna J M AJ   Kron Tomas T  

Journal of applied clinical medical physics 20130107 1


For the purpose of dose measurement using a high-dose rate (192)Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the (192)Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy  ...[more]

Similar Datasets

| S-EPMC5456930 | biostudies-other
| PRJEB29363 | ENA
2015-07-01 | GSE68008 | GEO
| S-EPMC2787446 | biostudies-literature
| S-EPMC6631330 | biostudies-literature
| S-EPMC8365654 | biostudies-literature
2015-07-01 | E-GEOD-68008 | biostudies-arrayexpress
| PRJEB11478 | ENA