Unknown

Dataset Information

0

In vivo proton dosimetry using a MOSFET detector in an anthropomorphic phantom with tissue inhomogeneity.


ABSTRACT: When in vivo proton dosimetry is performed with a metal-oxide semiconductor field-effect transistor (MOSFET) detector, the response of the detector depends strongly on the linear energy transfer. The present study reports a practical method to correct the MOSFET response for linear energy transfer dependence by using a simplified Monte Carlo dose calculation method (SMC). A depth-output curve for a mono-energetic proton beam in polyethylene was measured with the MOSFET detector. This curve was used to calculate MOSFET output distributions with the SMC (SMC(MOSFET)). The SMC(MOSFET) output value at an arbitrary point was compared with the value obtained by the conventional SMC(PPIC), which calculates proton dose distributions by using the depth-dose curve determined by a parallel-plate ionization chamber (PPIC). The ratio of the two values was used to calculate the correction factor of the MOSFET response at an arbitrary point. The dose obtained by the MOSFET detector was determined from the product of the correction factor and the MOSFET raw dose. When in vivo proton dosimetry was performed with the MOSFET detector in an anthropomorphic phantom, the corrected MOSFET doses agreed with the SMC(PPIC) results within the measurement error. To our knowledge, this is the first report of successful in vivo proton dosimetry with a MOSFET detector.

SUBMITTER: Kohno R 

PROVIDER: S-EPMC5716407 | biostudies-literature | 2012 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

In vivo proton dosimetry using a MOSFET detector in an anthropomorphic phantom with tissue inhomogeneity.

Kohno Ryosuke R   Hotta Kenji K   Matsubara Kana K   Nishioka Shie S   Matsuura Taeko T   Kawashima Mitsuhiko M  

Journal of applied clinical medical physics 20120308 2


When in vivo proton dosimetry is performed with a metal-oxide semiconductor field-effect transistor (MOSFET) detector, the response of the detector depends strongly on the linear energy transfer. The present study reports a practical method to correct the MOSFET response for linear energy transfer dependence by using a simplified Monte Carlo dose calculation method (SMC). A depth-output curve for a mono-energetic proton beam in polyethylene was measured with the MOSFET detector. This curve was u  ...[more]

Similar Datasets

| S-EPMC5718688 | biostudies-literature
| S-EPMC9388595 | biostudies-literature
| S-EPMC8885607 | biostudies-literature
| S-EPMC6865090 | biostudies-literature
| S-EPMC6871565 | biostudies-literature
| S-EPMC5690001 | biostudies-literature
| S-EPMC7428081 | biostudies-literature
| S-EPMC9042835 | biostudies-literature
| S-EPMC8080257 | biostudies-literature
| S-EPMC3371078 | biostudies-other