Kohno Ryosuke R Hotta Kenji K Matsubara Kana K Nishioka Shie S Matsuura Taeko T Kawashima Mitsuhiko M
Journal of applied clinical medical physics 20120308 2
When in vivo proton dosimetry is performed with a metal-oxide semiconductor field-effect transistor (MOSFET) detector, the response of the detector depends strongly on the linear energy transfer. The present study reports a practical method to correct the MOSFET response for linear energy transfer dependence by using a simplified Monte Carlo dose calculation method (SMC). A depth-output curve for a mono-energetic proton beam in polyethylene was measured with the MOSFET detector. This curve was u ...[more]