Comparison of Step-and-Shoot and Continuous-Bed-Motion PET Modes of Acquisition for Limited-View Organ Scans.


ABSTRACT: Continuous-bed-motion (CBM) acquisition mode has been made commercially available in PET/CT scanners. CBM mode is designed for whole-body imaging, with a long scan length (multiple axial fields of view [aFOVs]) and short acquisition duration (2-3 min/aFOV). PET/CT has recently been used after 90Y-microsphere therapy to quantify 90Y activity distribution in the liver. Here we compared counting efficiencies along the bed-motion direction (z-axis) between CBM and step-and-shoot (SS) acquisition modes for limited-view organ scans, such as 90Y PET/CT liver studies, that have short scan lengths (?2 aFOVs) and long acquisition durations (10-30 min/aFOV). Methods: The counting efficiencies, that is, analytic sensitivities, in SS mode (single-aFOV and multiple-aFOV scans) and CBM mode were theoretically derived and experimentally validated using a cylindric 68Ge phantom. The sensitivities along the z-axis were compared between the SS and CBM modes. Results: The analytic and experimental count profiles were in good agreement, validating the analytic models. For fixed scan durations, the overall coincidence counting efficiency in CBM mode was lower (?60%) than those in SS modes, and the maximum sensitivity in CBM mode was 50% or less of that in 1-aFOV SS mode and 100% or less of that in 2-aFOV SS mode. Conclusion: The ability of CBM mode to tailor-fit the PET/CT scan length and local scan duration is not realized in studies with a short scan length (?30 cm) and long scan duration (20 min/aFOV for the scanner). SS acquisition mode is preferable to CBM mode for limited-view organ and count-starved scans, such as 90Y PET/CT liver scans, because of the higher counting efficiency of SS mode, which leads to better image quality and quantification precision.

SUBMITTER: Siman W 

PROVIDER: S-EPMC5721725 | biostudies-literature | 2017 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of Step-and-Shoot and Continuous-Bed-Motion PET Modes of Acquisition for Limited-View Organ Scans.

Siman Wendy W   Kappadath S Cheenu SC  

Journal of nuclear medicine technology 20171017 4


Continuous-bed-motion (CBM) acquisition mode has been made commercially available in PET/CT scanners. CBM mode is designed for whole-body imaging, with a long scan length (multiple axial fields of view [aFOVs]) and short acquisition duration (2-3 min/aFOV). PET/CT has recently been used after <sup>90</sup>Y-microsphere therapy to quantify <sup>90</sup>Y activity distribution in the liver. Here we compared counting efficiencies along the bed-motion direction (<i>z</i>-axis) between CBM and step-a  ...[more]

Similar Datasets

| S-EPMC3843761 | biostudies-other
| 2252568 | ecrin-mdr-crc
| S-EPMC3871368 | biostudies-literature
| S-EPMC3285397 | biostudies-literature
| S-EPMC10202019 | biostudies-literature
| S-EPMC3269682 | biostudies-literature
| S-EPMC5857466 | biostudies-literature
| S-EPMC3806811 | biostudies-literature
| S-EPMC9339696 | biostudies-literature
| S-EPMC7648799 | biostudies-literature