A Randomized Crossover Trial of a Pressure Relief Technology (SensAwake™) in Continuous Positive Airway Pressure to Treat Obstructive Sleep Apnea.
Ontology highlight
ABSTRACT: Objectives/Background:Improving adherence to CPAP devices is crucial to reduce the long-term morbidity associated with OSA. SensAwake is a unique pressure relief technology that aims to promptly reduce the pressure upon sensing irregular respiration indicative of wakefulness. The purpose of this study was to compare adherence and sleep-quality outcomes in patients treated by CPAP with and without SensAwake technology. Methods:Participants with moderate-to-severe OSA were randomized to use CPAP devices with or without SensAwake (4 weeks) before crossing over. Results:Sixty-five patients completed both arms of the trial. There were no statistically significant differences in CPAP adherence with or without SensAwake over the study period (SensAwake ON 272.67 ± 17.06 versus SensAwake OFF 289.09 ± 15.24; p = 0.180). SensAwake reported a significantly lower system leak, 90th percentile leak, and time spent with excessive (>60?L/min) leak. Subgroup analysis suggested a trend towards improved adherence in patients with moderate-to-severe insomnia when using SensAwake. Conclusions:Using SensAwake incurred benefit in terms of reduced leaks; however, SensAwake did not improve CPAP adherence or objective sleep quality. Further studies should investigate the accuracy of observed trends towards increased adherence using SensAwake among patients with OSA and insomnia.