Unknown

Dataset Information

0

Torso geometry reconstruction and body surface electrode localization using three-dimensional photography.


ABSTRACT: We conducted a prospective clinical study (n=14; 29% female) to assess the accuracy of a three-dimensional (3D) photography-based method of torso geometry reconstruction and body surface electrodes localization. The position of 74 body surface electrocardiographic (ECG) electrodes (diameter 5mm) was defined by two methods: 3D photography, and CT (marker diameter 2mm) or MRI (marker size 10×20mm) imaging. Bland-Altman analysis showed good agreement in X (bias -2.5 [95% limits of agreement (LoA) -19.5 to 14.3] mm), Y (bias -0.1 [95% LoA -14.1 to 13.9] mm), and Z coordinates (bias -0.8 [95% LoA -15.6 to 14.2] mm), as defined by the CT/MRI imaging, and 3D photography. The average Hausdorff distance between the two torso geometry reconstructions was 11.17±3.05mm. Thus, accurate torso geometry reconstruction using 3D photography is feasible. Body surface ECG electrodes coordinates as defined by the CT/MRI imaging, and 3D photography, are in good agreement.

SUBMITTER: Perez-Alday EA 

PROVIDER: S-EPMC5776048 | biostudies-literature | 2018 Jan - Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Torso geometry reconstruction and body surface electrode localization using three-dimensional photography.

Perez-Alday Erick A EA   Thomas Jason A JA   Kabir Muammar M   Sedaghat Golriz G   Rogovoy Nichole N   van Dam Eelco E   van Dam Peter P   Woodward William W   Fuss Cristina C   Ferencik Maros M   Tereshchenko Larisa G LG  

Journal of electrocardiology 20170901 1


We conducted a prospective clinical study (n=14; 29% female) to assess the accuracy of a three-dimensional (3D) photography-based method of torso geometry reconstruction and body surface electrodes localization. The position of 74 body surface electrocardiographic (ECG) electrodes (diameter 5mm) was defined by two methods: 3D photography, and CT (marker diameter 2mm) or MRI (marker size 10×20mm) imaging. Bland-Altman analysis showed good agreement in X (bias -2.5 [95% limits of agreement (LoA) -  ...[more]

Similar Datasets

| S-EPMC3917702 | biostudies-literature
| S-EPMC7161442 | biostudies-literature
| S-EPMC3504357 | biostudies-literature
| S-EPMC3815983 | biostudies-literature
| S-EPMC7438683 | biostudies-literature
| S-EPMC4543434 | biostudies-literature
| S-EPMC3247630 | biostudies-literature