Perez-Alday Erick A EA Thomas Jason A JA Kabir Muammar M Sedaghat Golriz G Rogovoy Nichole N van Dam Eelco E van Dam Peter P Woodward William W Fuss Cristina C Ferencik Maros M Tereshchenko Larisa G LG
Journal of electrocardiology 20170901 1
We conducted a prospective clinical study (n=14; 29% female) to assess the accuracy of a three-dimensional (3D) photography-based method of torso geometry reconstruction and body surface electrodes localization. The position of 74 body surface electrocardiographic (ECG) electrodes (diameter 5mm) was defined by two methods: 3D photography, and CT (marker diameter 2mm) or MRI (marker size 10×20mm) imaging. Bland-Altman analysis showed good agreement in X (bias -2.5 [95% limits of agreement (LoA) - ...[more]