Difference between beta1-adrenoceptor autoantibodies of human and animal origin-Limitations detecting beta1-adrenoceptor autoantibodies using peptide based ELISA technology.
Ontology highlight
ABSTRACT: Cell-based analytics for the detection of the beta1-adrenoceptor autoantibody (beta1-AAB) are functional, yet difficult to handle, and should be replaced by easily applicable, routine lab methods. Endeavors to develop solid-phase-based assays such as ELISA to exploit epitope moieties for trapping autoantibodies are ongoing. These solid-phase-based assays, however, are often unreliable when used with human patient material, in contrast to animal derived autoantibodies. We therefore tested an immunogen peptide-based ELISA for the detection of beta1-AAB, and compared commercially available goat antibodies against the 2nd extracellular loop of human beta1-adrenoceptor (ADRB1-AB) to autoantibodies enriched from patient material. The functionality of these autoantibodies was tested in a cell based assay for comparison and their structural appearance was investigated using 2D gel electrophoresis. The ELISA showed a limit of detection for ADRB1-AB of about 1.5 nmol antibody/L when spiked in human control serum and only about 25 nmol/L when spiked in species identical (goat) matrix material. When applied to samples of human origin, the ELISA failed to identify the specific beta1-AABs. A low concentration of beta1-AAB, together with structural inconsistency of the patient originated samples as seen from the 2D Gel appearance, might contribute to the failure of the peptide based ELISA technology to detect human beta1-AABs.