Unknown

Dataset Information

0

Cardiac magnetic resonance T1 and extracellular volume mapping with motion correction and co-registration based on fast elastic image registration.


ABSTRACT:

Objective

Our aim was to investigate the technical feasibility of a novel motion compensation method for cardiac magntic resonance (MR) T1 and extracellular volume fraction (ECV) mapping.

Materials and methods

Native and post-contrast T1 maps were obtained using modified look-locker inversion recovery (MOLLI) pulse sequences with acquisition scheme defined in seconds. A nonrigid, nonparametric, fast elastic registration method was applied to generate motion-corrected T1 maps and subsequently ECV maps. Qualitative rating was performed based on T1 fitting-error maps and overlay images. Local deformation vector fields were produced for quantitative assessment. Intra- and inter-observer reproducibility were compared with and without motion compensation.

Results

Eighty-two T1 and 39 ECV maps were obtained in 21 patients with diverse myocardial diseases. Approximately 60% demonstrated clear quality improvement after motion correction for T1 mapping, particularly for the poor-rating cases (23% before vs 2% after). Approximately 67% showed further improvement with co-registration in ECV mapping. Although T1 and ECV values were not clinically significantly different before and after motion compensation, there was improved intra- and inter-observer reproducibility after motion compensation.

Conclusions

Automated motion correction and co-registration improved the qualitative assessment and reproducibility of cardiac MR T1 and ECV measurements, allowing for more reliable ECV mapping.

SUBMITTER: Zhang S 

PROVIDER: S-EPMC5813095 | biostudies-literature | 2018 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Cardiac magnetic resonance T1 and extracellular volume mapping with motion correction and co-registration based on fast elastic image registration.

Zhang Shuo S   Le Thu Thao TT   Kabus Sven S   Su Boyang B   Hausenloy Derek J DJ   Cook Stuart A SA   Chin Calvin W L CWL   Tan Ru San RS  

Magma (New York, N.Y.) 20171221 1


<h4>Objective</h4>Our aim was to investigate the technical feasibility of a novel motion compensation method for cardiac magntic resonance (MR) T1 and extracellular volume fraction (ECV) mapping.<h4>Materials and methods</h4>Native and post-contrast T1 maps were obtained using modified look-locker inversion recovery (MOLLI) pulse sequences with acquisition scheme defined in seconds. A nonrigid, nonparametric, fast elastic registration method was applied to generate motion-corrected T1 maps and s  ...[more]

Similar Datasets

| S-EPMC4158317 | biostudies-literature
| S-EPMC5882796 | biostudies-literature
| S-EPMC7116020 | biostudies-literature
| S-EPMC5820253 | biostudies-literature
| S-EPMC7688593 | biostudies-literature
| S-EPMC9992053 | biostudies-literature
| S-EPMC10113900 | biostudies-literature
| S-EPMC10388519 | biostudies-literature
| S-EPMC2700196 | biostudies-other
| S-EPMC4880324 | biostudies-literature