Qualification of NISTmAb charge heterogeneity control assays.
Ontology highlight
ABSTRACT: The NISTmAb is a monoclonal antibody Reference Material from the National Institute of Standards and Technology; it is a class-representative IgG1? intended serve as a pre-competitive platform for harmonization and technology development in the biopharmaceutical industry. The publication series of which this paper is a part describes NIST's overall control strategy to ensure NISTmAb quality and availability over its lifecycle. In this paper, the development and qualification of methods for monitoring NISTmAb charge heterogeneity are described. Capillary zone electrophoresis (CZE) and capillary isoelectric focusing (CIEF) assays were optimized and evaluated as candidate assays for NISTmAb quality control. CIEF was found to be suitable as a structural characterization assay yielding information on the apparent pI of the NISTmAb. CZE was found to be better suited for routine monitoring of NISTmAb charge heterogeneity and was qualified for this purpose. This paper is intended to provide relevant details of NIST's charge heterogeneity control strategy to facilitate implementation of the NISTmAb as a test molecule in the end user's laboratory. Graphical Abstract Representative capillary zone electropherogram of the NIST monoclonal antibody (NISTmAb). The NISTmAb is a publicly available research tool intended to facilitate advancement of biopharmaceutical analytics.