Unknown

Dataset Information

0

Performance of a new real-time continuous glucose monitoring system: A multicenter pilot study.


ABSTRACT:

Aims/introduction

The present study aimed to investigate the performance of a new real-time continuous glucose monitoring system.

Materials and methods

Interstitial glucose levels were monitored for 7 days in 63 patients with type 1 or type 2 diabetes using the Medtrum A6 TouchCare® CGM System. Venous blood was collected on a randomized day of the wear period. Plasma glucose levels were measured as reference values.

Results

Among 1,678 paired sensor-reference values, 90.5% (95% confidence interval 89.1-91.9%) were within ±20%/20 mg/dL of the reference values, with a mean absolute relative difference of 9.1 ± 8.7% (95% CI: 8.9-9.2%). The percentages of paired sensor-reference values falling within zone A and B of the Clarke error grid analysis (EGA) and the type 1 diabetes consensus EGA were 99.1 and 99.8%. Continuous EGA showed that the percentages of accurate readings, benign errors, and erroneous readings were 89.9, 6.3 and 3.8%, respectively. Surveillance EGA showed that 90.6, 9.2, and 0.2% of sensor-reference values with no, slight and lower moderate risk, respectively. The mean absolute relative difference was 16.6, and 96.0% of the sensor values fell within zones A and B of the consensus EGA for hypoglycemia. More than 85% of sensor values were within ±20%/20 mg/dL of reference values, the mean absolute relative difference was <11, and >99.5% of the sensor values fell in zones A and B of the consensus EGA.

Conclusions

The Medtrum real-time continuous glucose monitoring system was numerically and clinically accurate over a large glucose range across 7 days of wear.

SUBMITTER: Zhou J 

PROVIDER: S-EPMC5835467 | biostudies-literature | 2018 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance of a new real-time continuous glucose monitoring system: A multicenter pilot study.

Zhou Jian J   Zhang Shuo S   Li Liang L   Wang Yufei Y   Lu Wei W   Sheng Chunjun C   Li Yiming Y   Bao Yuqian Y   Jia Weiping W  

Journal of diabetes investigation 20170707 2


<h4>Aims/introduction</h4>The present study aimed to investigate the performance of a new real-time continuous glucose monitoring system.<h4>Materials and methods</h4>Interstitial glucose levels were monitored for 7 days in 63 patients with type 1 or type 2 diabetes using the Medtrum A6 TouchCare<sup>®</sup> CGM System. Venous blood was collected on a randomized day of the wear period. Plasma glucose levels were measured as reference values.<h4>Results</h4>Among 1,678 paired sensor-reference val  ...[more]

Similar Datasets

| S-EPMC6422005 | biostudies-literature
| S-EPMC6399785 | biostudies-literature
| S-EPMC4305078 | biostudies-literature
| S-EPMC7576426 | biostudies-literature
| S-EPMC2827490 | biostudies-literature
| S-EPMC4392936 | biostudies-literature
| S-EPMC10611957 | biostudies-literature
| S-EPMC5643124 | biostudies-literature
| S-EPMC8624400 | biostudies-literature
| S-EPMC8586329 | biostudies-literature