Human and model observer performance for lesion detection in breast cone beam CT images with the FDK reconstruction.


ABSTRACT: We investigate the detectability of breast cone beam computed tomography images using human and model observers and the variations of exponent, ?, of the inverse power-law spectrum for various reconstruction filters and interpolation methods in the Feldkamp-Davis-Kress (FDK) reconstruction. Using computer simulation, a breast volume with a 50% volume glandular fraction and a 2mm diameter lesion are generated and projection data are acquired. In the FDK reconstruction, projection data are apodized using one of three reconstruction filters; Hanning, Shepp-Logan, or Ram-Lak, and back-projection is performed with and without Fourier interpolation. We conduct signal-known-exactly and background-known-statistically detection tasks. Detectability is evaluated by human observers and their performance is compared with anthropomorphic model observers (a non-prewhitening observer with eye filter (NPWE) and a channelized Hotelling observer with either Gabor channels or dense difference-of-Gaussian channels). Our results show that the NPWE observer with a peak frequency of 7cyc/degree attains the best correlation with human observers for the various reconstruction filters and interpolation methods. We also discover that breast images with smaller ? do not yield higher detectability in the presence of quantum noise.

SUBMITTER: Han M 

PROVIDER: S-EPMC5854363 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

Human and model observer performance for lesion detection in breast cone beam CT images with the FDK reconstruction.

Han Minah M   Kim Byeongjoon B   Baek Jongduk J  

PloS one 20180315 3


We investigate the detectability of breast cone beam computed tomography images using human and model observers and the variations of exponent, β, of the inverse power-law spectrum for various reconstruction filters and interpolation methods in the Feldkamp-Davis-Kress (FDK) reconstruction. Using computer simulation, a breast volume with a 50% volume glandular fraction and a 2mm diameter lesion are generated and projection data are acquired. In the FDK reconstruction, projection data are apodize  ...[more]

Similar Datasets

| S-EPMC7067468 | biostudies-literature
| S-EPMC3598867 | biostudies-other
| S-EPMC5690196 | biostudies-literature
| S-EPMC4082266 | biostudies-other
| S-EPMC3828023 | biostudies-other
| S-EPMC3724792 | biostudies-other
| S-EPMC3069989 | biostudies-other
| S-EPMC3247929 | biostudies-literature
| S-EPMC3635775 | biostudies-other
| S-EPMC6739166 | biostudies-literature