Short term evaluation of respiratory effort by premature infants supported with bubble nasal continuous airway pressure using Seattle-PAP and a standard bubble device.


ABSTRACT:

Background

Almost one million prematurely born infants die annually from respiratory insufficiency, predominantly in countries with limited access to respiratory support for neonates. The primary hypothesis tested in the present study was that a modified device for bubble nasal continuous positive airway pressure (Bn-CPAP) would provide lower work of spontaneous breathing, estimated by esophageal pressure-rate products.

Methods

Infants born <32 weeks gestation and stable on Bn-CPAP with FiO2 <0.30 were studied within 72 h following delivery. Esophageal pressures during spontaneous breathing were measured during 2 h on standard Bn-CPAP, then 2 h with Bn-CPAP using a modified bubble device presently termed Seattle-PAP, which produces a different pattern of pressure fluctuation