Unknown

Dataset Information

0

Accuracy and practicality of a portable ozone monitor for personal exposure estimates.


ABSTRACT: Accurate measurements of personal exposure to atmospheric pollutants such as ozone are important for understanding health risks. We tested a new personal ozone monitor (POM; 2B Technologies) for accuracy, precision, and ease of use. The POM's measurements were compared to simultaneous ozone measurements from a 2B Model 205 monitor and a ThermoScientific 49i monitor, and multiple POMs were placed side-by-side to check precision. Tests were undertaken in a controlled environmental facility, outdoors, and in a private residence. Additionally, ten volunteers wore a POM for five days and answered a questionnaire about its ease of use. The POM measured ozone accurately compared to the 49i ozone monitor, with average relative differences of less than 8%. In the controlled environment tests, the POM's ozone measurements did not change in the presence of additional atmospheric constituents with similar absorption lines to ozone, though there may have been a small decrease in precision and accuracy. Precision between POMs varied by environment (r2 = 0.98 outdoors; r2 = 0.3 to 0.9 in controlled lab conditions). Volunteers reported that the POM was reasonably comfortable to wear, although all reported that they felt that it was too noisy. Overall, the POM is a viable option for personal ozone monitoring.

SUBMITTER: Sagona JA 

PROVIDER: S-EPMC5918273 | biostudies-literature | 2018 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy and practicality of a portable ozone monitor for personal exposure estimates.

Sagona Jessica A JA   Weisel Clifford C   Meng Qingyu Q  

Atmospheric environment (Oxford, England : 1994) 20171124


Accurate measurements of personal exposure to atmospheric pollutants such as ozone are important for understanding health risks. We tested a new personal ozone monitor (POM; 2B Technologies) for accuracy, precision, and ease of use. The POM's measurements were compared to simultaneous ozone measurements from a 2B Model 205 monitor and a ThermoScientific 49i monitor, and multiple POMs were placed side-by-side to check precision. Tests were undertaken in a controlled environmental facility, outdoo  ...[more]

Similar Datasets

| S-EPMC7273279 | biostudies-literature
| S-EPMC8314631 | biostudies-literature
| S-EPMC6751078 | biostudies-literature
| S-EPMC5263080 | biostudies-other
| S-EPMC6164905 | biostudies-literature
| S-EPMC6050681 | biostudies-literature
| S-EPMC8318339 | biostudies-literature
| S-EPMC5880233 | biostudies-literature
| S-EPMC6330888 | biostudies-literature
| S-EPMC6535137 | biostudies-literature