Aortic root evaluation prior to transcatheter aortic valve implantation-Correlation of manual and semi-automatic measurements.
Ontology highlight
ABSTRACT: BACKGROUND:Pre-procedural TAVI planning requires highly sophisticated and time-consuming manual measurements performed by experienced readers. Semi-automatic software may assist with partial automation of assessment of multiple parameters. The aim of this study was to evaluate differences between manual and semi-automatic measurements in terms of agreement and time. METHODS:One hundred and twenty TAVI candidates referred for the retrospectively ECG-gated CTA (2nd and 3rd generation dual source CT) were evaluated. Fully manual and semi-automatic measurements of fourteen aortic root parameters were assessed in the 20% phase of the R-R interval. Reading time was compared using paired samples t-test. Inter-software agreement was calculated using the Intraclass correlation coefficient (ICC) in a 2-way mixed effects model. Differences between manual and semi-automatic measurements were evaluated using Bland-Altman analysis. RESULTS:The time needed for evaluation using semi-automatic assessment (3 min 24 s ± 1 min 7 s) was significantly lower (p<0.001) compared to a fully manual approach (6 min 31 sec ± 1 min 1 sec). Excellent inter-software agreement was found (ICC = 0.93 ± 0.0; range:0.90-0.95). The same prosthesis size from manual and semi-automatic measurements was selected in 92% of cases, when sizing was based on annular area. Prosthesis sizing based on annular short diameter and perimeter agreed in 99% and 96% cases, respectively. CONCLUSION:Use of semi-automatic software in pre-TAVI evaluation results in comparable results in respect of measurements and selected valve prosthesis size, while necessary reading time is significantly lower.
SUBMITTER: Horehledova B
PROVIDER: S-EPMC6023104 | biostudies-literature | 2018
REPOSITORIES: biostudies-literature
ACCESS DATA