The residual STL volume as a metric to evaluate accuracy and reproducibility of anatomic models for 3D printing: application in the validation of 3D-printable models of maxillofacial bone from reduced radiation dose CT images.


ABSTRACT: Background:The effects of reduced radiation dose CT for the generation of maxillofacial bone STL models for 3D printing is currently unknown. Images of two full-face transplantation patients scanned with non-contrast 320-detector row CT were reconstructed at fractions of the acquisition radiation dose using noise simulation software and both filtered back-projection (FBP) and Adaptive Iterative Dose Reduction 3D (AIDR3D). The maxillofacial bone STL model segmented with thresholding from AIDR3D images at 100 % dose was considered the reference. For all other dose/reconstruction method combinations, a "residual STL volume" was calculated as the topologic subtraction of the STL model derived from that dataset from the reference and correlated to radiation dose. Results:The residual volume decreased with increasing radiation dose and was lower for AIDR3D compared to FBP reconstructions at all doses. As a fraction of the reference STL volume, the residual volume decreased from 2.9 % (20 % dose) to 1.4 % (50 % dose) in patient 1, and from 4.1 % to 1.9 %, respectively in patient 2 for AIDR3D reconstructions. For FBP reconstructions it decreased from 3.3 % (20 % dose) to 1.0 % (100 % dose) in patient 1, and from 5.5 % to 1.6 %, respectively in patient 2. Its morphology resembled a thin shell on the osseous surface with average thickness <0.1 mm. Conclusion:The residual volume, a topological difference metric of STL models of tissue depicted in DICOM images supports that reduction of CT dose by up to 80 % of the clinical acquisition in conjunction with iterative reconstruction yields maxillofacial bone models accurate for 3D printing.

SUBMITTER: Cai T 

PROVIDER: S-EPMC6036610 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

The residual STL volume as a metric to evaluate accuracy and reproducibility of anatomic models for 3D printing: application in the validation of 3D-printable models of maxillofacial bone from reduced radiation dose CT images.

Cai Tianrun T   Rybicki Frank J FJ   Giannopoulos Andreas A AA   Schultz Kurt K   Kumamaru Kanako K KK   Liacouras Peter P   Demehri Shadpour S   Shu Small Kirstin M KM   Mitsouras Dimitris D  

3D printing in medicine 20151127 1


<h4>Background</h4>The effects of reduced radiation dose CT for the generation of maxillofacial bone STL models for 3D printing is currently unknown. Images of two full-face transplantation patients scanned with non-contrast 320-detector row CT were reconstructed at fractions of the acquisition radiation dose using noise simulation software and both filtered back-projection (FBP) and Adaptive Iterative Dose Reduction 3D (AIDR3D). The maxillofacial bone STL model segmented with thresholding from  ...[more]

Similar Datasets

| S-EPMC5915392 | biostudies-literature
| S-EPMC6604582 | biostudies-literature
| S-EPMC4495599 | biostudies-literature
| S-EPMC5408980 | biostudies-literature
| S-EPMC5573584 | biostudies-other
| S-EPMC6969138 | biostudies-literature
| S-EPMC7417997 | biostudies-literature
| S-EPMC8362236 | biostudies-literature
| S-EPMC7317697 | biostudies-literature
| S-EPMC8127415 | biostudies-literature