Rauwolf M M Turyanskaya A A Ingerle D D Szoboszlai N N Pape I I Malandain A W AW Fox O J L OJL Hahn L L Sawhney K J S KJS Streli C C
Journal of synchrotron radiation 20180613 Pt 4
An X-ray fluorescence setup has been tested on the B16 beamline at the Diamond Light Source synchrotron with two different excitation energies (12.7 and 17 keV). This setup allows the scanning of thin samples (thicknesses up to several micrometers) with a sub-micrometer resolution (beam size of 500 nm × 600 nm determined with a 50 µm Au wire). Sensitivities and detection limits reaching values of 249 counts s<sup>-1</sup> fg<sup>-1</sup> and 4 ag in 1000 s, respectively (for As Kα excited with 1 ...[more]