Unknown

Dataset Information

0

Validation of Miniaturized Particulate-Matter Real-Time Samplers for Characterizing Personal Polycyclic Aromatic Hydrocarbon Exposure.


ABSTRACT: This study validates the analysis of polycyclic aromatic hydrocarbons (PAHs) in microgram levels of particulate matter (PM) collected on filters by two low-flow rate, real-time monitors, microPEM™ and microAeth®. Particle-associated PAHs were analyzed by a coupling of a gas chromatograph to a sensitive, atmospheric-pressure laser ionization-mass spectrometer. Air particulate samples were collected over the course of one or two days in the living room of a fourth-floor apartment in New York City. Three types of samplers, the two aforementioned personal samplers and a high-flow rate pump (4 liters per minute), were operated side by side, and three samples of each type were collected during each sampling period. Intrasampler agreement as measured by relative standard deviation (RSD) was within 1% to 18%. After background subtraction, total PAH measured by all three sampler types had good agreement (R=0.99). This ability to accurately characterize personal PAH exposure in archived filters collected by these real-time samplers could provide additional important PAH exposure information that can benefit many environmental health studies using these monitors.

SUBMITTER: Yan B 

PROVIDER: S-EPMC6039126 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of Miniaturized Particulate-Matter Real-Time Samplers for Characterizing Personal Polycyclic Aromatic Hydrocarbon Exposure.

Yan Beizhan B   Pitiranggon Masha M   Ross James J   Arthen-Engeland Thomas T   Stelter Andreas A   Chillrud Steven N SN  

Journal of analytical & bioanalytical techniques 20180411 2


This study validates the analysis of polycyclic aromatic hydrocarbons (PAHs) in microgram levels of particulate matter (PM) collected on filters by two low-flow rate, real-time monitors, microPEM™ and microAeth<sup>®</sup>. Particle-associated PAHs were analyzed by a coupling of a gas chromatograph to a sensitive, atmospheric-pressure laser ionization-mass spectrometer. Air particulate samples were collected over the course of one or two days in the living room of a fourth-floor apartment in New  ...[more]

Similar Datasets

| S-EPMC5577249 | biostudies-literature
| S-EPMC4046869 | biostudies-other
| S-EPMC2914693 | biostudies-literature
| S-EPMC3908783 | biostudies-literature
| S-EPMC8175174 | biostudies-literature
| S-EPMC4829995 | biostudies-literature
| S-EPMC205371 | biostudies-other
| S-EPMC3436775 | biostudies-literature
| PRJNA793526 | ENA
| S-EPMC6972610 | biostudies-literature