A novel sideways fall simulator to study hip fractures ex vivo.


ABSTRACT: Falls to the side are the leading cause of hip fractures in the elderly. The load that a person experiences during a fall cannot be measured with volunteers for ethical reasons. To evaluate injurious loads, while considering relevant energy input and body posture for a sideways fall, a subject-specific cadaveric impact experiment was developed. Full cadaveric femur-pelvis constructs (N = 2) were embedded in surrogate soft tissue material and attached to metallic surrogate lower limbs. The specimens were then subjected to an inverted pendulum motion, simulating a fall to the side with an impact to the greater trochanter. The load at the ground and the deformation of the pelvis were evaluated using a 6-axis force transducer and two high-speed cameras. Post-test, a trauma surgeon (PG) evaluated specimen injuries. Peak ground contact forces were 7132 N and 5641 N for the fractured and non-fractured specimen, respectively. We observed a cervical fracture of the femur in one specimen and no injuries in a second specimen, showing that the developed protocol can be used to differentiate between specimens at high and low fracture risk.

SUBMITTER: Fleps I 

PROVIDER: S-EPMC6057661 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

A novel sideways fall simulator to study hip fractures ex vivo.

Fleps Ingmar I   Vuille Muriel M   Melnyk Angela A   Ferguson Stephen J SJ   Guy Pierre P   Helgason Benedikt B   Cripton Peter A PA  

PloS one 20180724 7


Falls to the side are the leading cause of hip fractures in the elderly. The load that a person experiences during a fall cannot be measured with volunteers for ethical reasons. To evaluate injurious loads, while considering relevant energy input and body posture for a sideways fall, a subject-specific cadaveric impact experiment was developed. Full cadaveric femur-pelvis constructs (N = 2) were embedded in surrogate soft tissue material and attached to metallic surrogate lower limbs. The specim  ...[more]

Similar Datasets

| S-EPMC6095517 | biostudies-literature
| S-EPMC7769867 | biostudies-literature
| S-EPMC8362820 | biostudies-literature
| S-EPMC6738497 | biostudies-literature
| S-EPMC7804395 | biostudies-literature
| S-EPMC10865533 | biostudies-literature
| S-EPMC11364790 | biostudies-literature
| S-EPMC6754348 | biostudies-literature
| S-EPMC8390060 | biostudies-literature
| S-EPMC9076424 | biostudies-literature