Ontology highlight
ABSTRACT: Purpose
To compare aqueous humour (AH) dynamics in the presence of a precrystalline (Implantable Collamer Lens®; ICL) or iris-fixed (Artiflex®) phakic intraocular lens (PIOL).Methods
By computational fluid dynamics simulation, AH flow was modelled through a peripheral iridotomy (PI) or central lens hole (both 360 ?m) in the presence of an Artiflex or ICL lens, respectively. The impacts of AH flow were then determined in terms of wall shear stress (WSS) produced on the endothelium or crystalline lens. Effects were also modelled for different scenarios of pupil diameter (PD 3.5 or 5.5 mm), ICL vault (100, 350, 800 ?m) and number of Artiflex iridotomies (1 or 2) and location (12 or 6 o'clock).Results
For a PD of 3.5 mm, AH volumes flowing from the posterior to the anterior chamber were 37.6% of total flow through the lens hole (ICL) and 84.2% through PI (Artiflex). For an enlarged PD (5.5 mm), corresponding values were 10.3% and 81.9% respectively, so PI constitutes a very efficient way of evacuating AH. Central endothelial WSS in Pa was lower for the large vault ICL and the Artiflex (1-03 and 1.1-03 respectively) compared to the PIOL-free eye (1.6-03). Crystalline lens WSS was highest for the lowest vault ICL (1-04).Conclusions
AH flow varied according to the presence of a precrystalline or iris-fixed intraocular lens. Endothelial WSS was lower for an implanted ICL with large vault and Artiflex than in the PIOL-free eye, while highest crystalline WSS was recorded for the lowest vault ICL.
SUBMITTER: Fernandez-Vigo JI
PROVIDER: S-EPMC6089426 | biostudies-literature | 2018
REPOSITORIES: biostudies-literature
Fernández-Vigo José Ignacio JI Marcos Alfonso C AC Agujetas Rafael R Montanero José María JM Sánchez-Guillén Inés I García-Feijóo Julián J Pandal-Blanco Adrián A Fernández-Vigo José Ángel JÁ Macarro-Merino Ana A
PloS one 20180813 8
<h4>Purpose</h4>To compare aqueous humour (AH) dynamics in the presence of a precrystalline (Implantable Collamer Lens®; ICL) or iris-fixed (Artiflex®) phakic intraocular lens (PIOL).<h4>Methods</h4>By computational fluid dynamics simulation, AH flow was modelled through a peripheral iridotomy (PI) or central lens hole (both 360 μm) in the presence of an Artiflex or ICL lens, respectively. The impacts of AH flow were then determined in terms of wall shear stress (WSS) produced on the endothelium ...[more]