Unknown

Dataset Information

Quantitation of tadalafil in human plasma using a sensitive and rapid LC-MS/MS method for a bioequivalence study.


ABSTRACT: A highly sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed for the determination of tadalafil (TAD) in human plasma. TAD and its deuterated internal standard (IS), tadalafil-d3, were extracted from 200 µL plasma using Phenomenex Strata-X-C 33 µ extraction cartridges. Chromatographic analysis was carried out on Synergi™ Hydro-RP C18 (100 mm × 4.6 mm, 4 µm) column with a mobile phase consisting of methanol and 10 mM ammonium formate, pH 4.0 (90:10, v/v), delivered at a flow rate of 0.9 mL/min. Quantitation of the protonated analyte was done on a triple quadrupole mass spectrometer using multiple reaction monitoring via electrospray ionization. The precursor to product ions transitions monitored for TAD and TAD-d3 were m/z 390.3 → 268.2 and m/z 393.1 → 2

SUBMITTER: Bhadoriya A 

PROVIDER: S-EPMC6104147 | biostudies-literature | 2018 Aug

REPOSITORIES: biostudies-literature