Feasibility of A Novel On-Site Detection Method for Aflatoxin in Maize Flour from Markets and Selected Households in Kampala, Uganda.


ABSTRACT: In sub-Saharan Africa, there is a high demand for affordable and accessible methods for on-site detection of aflatoxins for appropriate food safety management. In this study, we validated an electrochemical immunosensor device by the on-site detection of 60 maize flour samples from six markets and 72 samples from households in Kampala. The immunosensor was successfully validated with a linear range from 0.7 ± 0.1 to 11 ± 0.3 µg/kg and limit of detection (LOD) of 0.7 µg/kg. The maize flour samples from the markets had a mean total aflatoxin concentration of 7.6 ± 2.3 µg/kg with approximately 20% of the samples higher than 10 µg/kg, which is the maximum acceptable level in East Africa. Further down the distribution chain, at the household level, approximately 45% of the total number contained total aflatoxin levels higher than the acceptable limit. The on-site detection method correlated well with the established laboratory-based HPLC and ELISA-detection methods for aflatoxin B? with the correlation coefficients of 0.94 and 0.98, respectively. This study shows the feasibility of a novel on-site detection method and articulates the severity of aflatoxin contamination in Uganda.

SUBMITTER: Paul Wacoo A 

PROVIDER: S-EPMC6115733 | biostudies-literature | 2018 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Feasibility of A Novel On-Site Detection Method for Aflatoxin in Maize Flour from Markets and Selected Households in Kampala, Uganda.

Paul Wacoo Alex A   Wendiro Deborah D   Nanyonga Sarah S   Hawumba Joseph F JF   Sybesma Wilbert W   Kort Remco R  

Toxins 20180811 8


In sub-Saharan Africa, there is a high demand for affordable and accessible methods for on-site detection of aflatoxins for appropriate food safety management. In this study, we validated an electrochemical immunosensor device by the on-site detection of 60 maize flour samples from six markets and 72 samples from households in Kampala. The immunosensor was successfully validated with a linear range from 0.7 ± 0.1 to 11 ± 0.3 µg/kg and limit of detection (LOD) of 0.7 µg/kg. The maize flour sample  ...[more]

Similar Datasets

| S-EPMC3038854 | biostudies-literature
| S-EPMC7695076 | biostudies-literature
| S-EPMC5443453 | biostudies-literature
| S-EPMC6456174 | biostudies-literature
| S-EPMC6874685 | biostudies-literature
| S-EPMC8058529 | biostudies-literature
| S-EPMC2394214 | biostudies-other