Effectiveness of skull X-RAY to determine cochlear implant insertion depth.


ABSTRACT:

Background

Cochlear implant (CI) insertion depth can affect residual hearing preservation, tonotopic range coverage, and Mapping. Therefore, determining insertion depth has the potential to maximize CI performance. A post-op skull X-RAY is commonly used to assess insertion depth, however its effectiveness has not been well established. Our primary objective was to assess the accuracy of post-op skull X-RAYs to determine insertion depth, compared to CT as the gold standard. Secondary objectives were to compare experience level of raters and different skull X-RAY views.

Methods

Thirteen patients with Advanced Bionic HiRes 90 K implants, and post-operative temporal bone CT scans were selected from the CI database at Sunnybrook Health Sciences Centre. Medical students, otology fellows, and CI surgeons evaluated insertion depths on post-op skull X-RAYs, while neuroradiologists evaluated CT scans. Descriptive statistics, regression analysis, and paired t-tests were used to compare the two types of imaging.

Results

X-RAYs and CTs provided an equivalent mean insertion depth of 337 degrees (p?=?0.93), a mean difference of -?0.9 degrees and a standard deviation of paired differences of 43 degrees. Although means were similar across rater groups, CI surgeons (45 degrees) had the lowest standard deviation of paired differences. Comparing X-RAY views, Caldwell (29 degrees) had less variation than Towne (59 degrees) for standard deviation of paired differences.

Conclusions

Skull X-RAYs provide accurate and reliable measurements for CI insertion depth. The Caldwell view alone may be sufficient for evaluations of insertion depth, and experience has a minor impact on the variability of estimates.

SUBMITTER: Fernandes V 

PROVIDER: S-EPMC6122652 | biostudies-literature | 2018 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Effectiveness of skull X-RAY to determine cochlear implant insertion depth.

Fernandes Vinay V   Wang Yiqiao Y   Yeung Robert R   Symons Sean S   Lin Vincent V  

Journal of otolaryngology - head & neck surgery = Le Journal d'oto-rhino-laryngologie et de chirurgie cervico-faciale 20180903 1


<h4>Background</h4>Cochlear implant (CI) insertion depth can affect residual hearing preservation, tonotopic range coverage, and Mapping. Therefore, determining insertion depth has the potential to maximize CI performance. A post-op skull X-RAY is commonly used to assess insertion depth, however its effectiveness has not been well established. Our primary objective was to assess the accuracy of post-op skull X-RAYs to determine insertion depth, compared to CT as the gold standard. Secondary obje  ...[more]

Similar Datasets

| S-EPMC7039896 | biostudies-literature
| S-EPMC6641467 | biostudies-literature
| S-EPMC6344356 | biostudies-literature
| S-EPMC2538415 | biostudies-other
| S-EPMC6435875 | biostudies-literature
| S-EPMC5763516 | biostudies-literature
| S-EPMC3872819 | biostudies-literature
| S-EPMC6671933 | biostudies-literature
| S-EPMC8488513 | biostudies-literature
| S-EPMC5335919 | biostudies-literature