Hyphenation of Supercritical Fluid Chromatography and NMR with In-Line Sample Concentration.


ABSTRACT: By coupling supercritical fluid chromatography (SFC) and nuclear magnetic resonance (NMR) in-line, a powerful analytical method arises that enables chemically specific analysis of a broad range of complex mixtures. However, during chromatography, the compounds are diluted in the mobile phase, in this case supercritical CO2 (scCO2), often resulting in concentrations that are too low to be detected by NMR spectroscopy or at least requiring excessive signal averaging. We present a hyphenated SFC-NMR setup with an integrated approach for concentrating samples in-line, which are diluted in scCO2 during chromatography. This in-line concentration is achieved by controlled in-line expansion of the scCO2. As a proof of concept four isomers of vitamin E (tocopherol) were isolated by SFC, concentrated in-line by expanding CO2 from 120 to 50 bar, and finally shuttled to the NMR spectrometer fitted with a dedicated probehead for spectroscopic characterization of microfluidic samples. The abundant isomers were readily detected, supporting the viability of SFC-NMR as a powerful analytical tool.

SUBMITTER: van Zelst FHM 

PROVIDER: S-EPMC6127797 | biostudies-literature | 2018 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Hyphenation of Supercritical Fluid Chromatography and NMR with In-Line Sample Concentration.

van Zelst F H M FHM   van Meerten S G J SGJ   van Bentum P J M PJM   Kentgens A P M APM  

Analytical chemistry 20180820 17


By coupling supercritical fluid chromatography (SFC) and nuclear magnetic resonance (NMR) in-line, a powerful analytical method arises that enables chemically specific analysis of a broad range of complex mixtures. However, during chromatography, the compounds are diluted in the mobile phase, in this case supercritical CO<sub>2</sub> (scCO<sub>2</sub>), often resulting in concentrations that are too low to be detected by NMR spectroscopy or at least requiring excessive signal averaging. We prese  ...[more]

Similar Datasets

| S-EPMC4522036 | biostudies-literature
| S-EPMC5662104 | biostudies-literature
| S-EPMC6193530 | biostudies-literature
| S-EPMC6972688 | biostudies-literature