Dog-bone inflation to salvage and redeploy a deformed, dislodged expandable stent in the external iliac artery.


ABSTRACT: A 64-year-old man with a history of right external iliac artery stenosis after stenting with a self-expandable stent was admitted for repetitive in-stent restenosis (ISR). To treat ISR, we selected a balloon expandable stent (BES) that was prepared with negative suction before being inserted into the sheath. BES dislodged when the stent balloon was incompletely inflated and was simultaneously deformed in a funnel shape. When BES was retrieved using a short noncompliant balloon, it dislodged again because of an imbalanced proximal inflation of the balloon. We finally retrieved BES using a long noncompliant balloon and deployed it with dog-bone inflation without any additional devices. <Learning objective: To report an easy method for retrieving and redeploying a deformed, dislodged stent without using an additional device. Retrieving a relatively large peripheral stent that becomes deformed and dislodged is difficult and may require surgery. Balloon catheter retrieval is a useful strategy to salvage and redeploy the dislodged stent. However, a large size short noncompliant balloon catheter had the risk for dislodging the retrieved deformed stent again. To retrieve and salvage a deformed and dislodged stent, a long balloon and dog-bone inflation may be useful.>.

SUBMITTER: Chung WJ 

PROVIDER: S-EPMC6149551 | biostudies-literature | 2018 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Dog-bone inflation to salvage and redeploy a deformed, dislodged expandable stent in the external iliac artery.

Chung Wen-Jung WJ   Takeji Yasuaki Y   Soga Yoshimitsu Y   Ando Kenji K  

Journal of cardiology cases 20171120 1


A 64-year-old man with a history of right external iliac artery stenosis after stenting with a self-expandable stent was admitted for repetitive in-stent restenosis (ISR). To treat ISR, we selected a balloon expandable stent (BES) that was prepared with negative suction before being inserted into the sheath. BES dislodged when the stent balloon was incompletely inflated and was simultaneously deformed in a funnel shape. When BES was retrieved using a short noncompliant balloon, it dislodged agai  ...[more]

Similar Datasets

| S-EPMC4920989 | biostudies-literature
| S-EPMC10562087 | biostudies-literature
| S-EPMC9349184 | biostudies-literature
| S-EPMC7771483 | biostudies-literature
| S-EPMC3331627 | biostudies-other
| S-EPMC4996068 | biostudies-literature
| S-EPMC9748016 | biostudies-literature
| S-EPMC5483202 | biostudies-other
| S-EPMC9059807 | biostudies-literature