Unknown

Dataset Information

0

Challenges and Complication Management in Novel Artificial Iris Implantation.


ABSTRACT:

Purpose

Evaluation of postoperative artificial iris prosthesis-related complications.

Design

Retrospective cohort study.

Methods

Fifty-one consecutive patients underwent pupillary reconstruction using an artificial iris implant made from silicone between 2011 and 2015. Quantity and quality of complications were subclassified into three groups including mild, moderate, and severe complications. Their management and the learning curve were evaluated.

Results

In total, 13 (25.5%) of 51 included artificial iris implantations showed unexpected events in various degrees: mild complications: recurrent bleeding (n=1, 2.0%), slight but stable iris deviation (n=1, 2.0%), capsular fibrosis (n=2, 3.9%); moderate complications: suture cutting through the residual iris (n=1, 2.0%), glaucoma (n=3, 5.9%), and corneal decompensation (n=3, 5.9%); severe complications: artificial iris suture loosening (n=2, 3.9%) and dislocation (n=3, 5.9%), synechiae (n=2, 3.9%), glaucoma (n=2, 3.9%), and corneal decompensation (n=5, 9.8%) with the need for surgery, cystoid macular edema (n=3, 5.9%) and retinal detachment (n=1, 2.0%). The complication rate decreased from 83.3% (5 of 6 implantations) in the first year to 13.3% (2 of 15 implantations) in the 4th year. Nineteen of 45 evaluated patients showed a significant gain in best-corrected visual acuity (BCVA) from 1.09?±?0.56 logMAR to 0.54?±?0.48 logMAR (p < 0.001), and 13 of 45 eyes had a significant BCVA loss from 0.48?±?0.39 logMAR to 0.93?±?0.41 logMAR after surgery (p < 0.001).

Conclusions

The artificial iris is a feasible option in the treatment of iris defects with a wide spectrum of postoperative complications. The significant reduction of complications after twelve implantations implicates that the procedure is not to be recommended in low volume settings.

SUBMITTER: Mayer CS 

PROVIDER: S-EPMC6174745 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

Challenges and Complication Management in Novel Artificial Iris Implantation.

Mayer Christian S CS   Laubichler Andrea E AE   Khoramnia Ramin R   Tandogan Tamer T   Prahs Philipp P   Zapp Daniel D   Reznicek Lukas L  

Journal of ophthalmology 20180923


<h4>Purpose</h4>Evaluation of postoperative artificial iris prosthesis-related complications.<h4>Design</h4>Retrospective cohort study.<h4>Methods</h4>Fifty-one consecutive patients underwent pupillary reconstruction using an artificial iris implant made from silicone between 2011 and 2015. Quantity and quality of complications were subclassified into three groups including mild, moderate, and severe complications. Their management and the learning curve were evaluated.<h4>Results</h4>In total,  ...[more]

Similar Datasets

| S-EPMC8121880 | biostudies-literature
| S-EPMC8443204 | biostudies-literature
| S-EPMC9297097 | biostudies-literature
| S-EPMC5989510 | biostudies-literature
| S-EPMC10251676 | biostudies-literature
| S-EPMC10498187 | biostudies-literature
| S-EPMC8311180 | biostudies-literature
| S-EPMC9175140 | biostudies-literature
| S-EPMC7474060 | biostudies-literature