Gradient nonlinearity effects on upper cervical spinal cord area measurement from 3D T1 -weighted brain MRI acquisitions.


ABSTRACT: PURPOSE:To explore (i) the variability of upper cervical cord area (UCCA) measurements from volumetric brain 3D T1 -weighted scans related to gradient nonlinearity (GNL) and subject positioning; (ii) the effect of vendor-implemented GNL corrections; and (iii) easily applicable methods that can be used to retrospectively correct data. METHODS:A multiple sclerosis patient was scanned at seven sites using 3T MRI scanners with the same 3D T1 -weighted protocol without GNL-distortion correction. Two healthy subjects and a phantom were additionally scanned at a single site with varying table positions. The 2D and 3D vendor-implemented GNL-correction algorithms and retrospective methods based on (i) phantom data fit, (ii) normalization with C2 vertebral body diameters, and (iii) the Jacobian determinant of nonlinear registrations to a template were tested. RESULTS:Depending on the positioning of the subject, GNL introduced up to 15% variability in UCCA measurements from volumetric brain T1 -weighted scans when no distortion corrections were used. The 3D vendor-implemented correction methods and the three proposed methods reduced this variability to less than 3%. CONCLUSIONS:Our results raise awareness of the significant impact that GNL can have on quantitative UCCA studies, and point the way to prospectively and retrospectively managing GNL distortions in a variety of settings, including clinical environments. Magn Reson Med 79:1595-1601, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

SUBMITTER: Papinutto N 

PROVIDER: S-EPMC6175276 | biostudies-literature | 2018 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Purpose</h4>To explore (i) the variability of upper cervical cord area (UCCA) measurements from volumetric brain 3D T<sub>1</sub> -weighted scans related to gradient nonlinearity (GNL) and subject positioning; (ii) the effect of vendor-implemented GNL corrections; and (iii) easily applicable methods that can be used to retrospectively correct data.<h4>Methods</h4>A multiple sclerosis patient was scanned at seven sites using 3T MRI scanners with the same 3D T<sub>1</sub> -weighted protocol wi  ...[more]

Similar Datasets

| S-EPMC8596767 | biostudies-literature
| S-EPMC6704046 | biostudies-literature
| S-EPMC5241082 | biostudies-literature
| S-EPMC9006307 | biostudies-literature
| S-EPMC5023083 | biostudies-literature
| S-EPMC7965580 | biostudies-literature
| S-EPMC10236479 | biostudies-literature
| S-EPMC10291889 | biostudies-literature
| S-EPMC3442779 | biostudies-literature
| S-EPMC7815834 | biostudies-literature