ABSTRACT: Purpose
To explore (i) the variability of upper cervical cord area (UCCA) measurements from volumetric brain 3D T1 -weighted scans related to gradient nonlinearity (GNL) and subject positioning; (ii) the effect of vendor-implemented GNL corrections; and (iii) easily applicable methods that can be used to retrospectively correct data.Methods
A multiple sclerosis patient was scanned at seven sites using 3T MRI scanners with the same 3D T1 -weighted protocol without GNL-distortion correction. Two healthy subjects and a phantom were additionally scanned at a single site with varying table positions. The 2D and 3D vendor-implemented GNL-correction algorithms and retrospective methods based on (i) phantom data fit, (ii) normalization with C2 vertebral body diameters, and (iii) the Jacobian determinant of nonlinear registrations to a template were tested.Results
Depending on the positioning of the subject, GNL introduced up to 15% variability in UCCA measurements from volumetric brain T1 -weighted scans when no distortion corrections were used. The 3D vendor-implemented correction methods and the three proposed methods reduced this variability to less than 3%.Conclusions
Our results raise awareness of the significant impact that GNL can have on quantitative UCCA studies, and point the way to prospectively and retrospectively managing GNL distortions in a variety of settings, including clinical environments. Magn Reson Med 79:1595-1601, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
SUBMITTER: Papinutto N
PROVIDER: S-EPMC6175276 | biostudies-literature | 2018 Mar
REPOSITORIES: biostudies-literature
Papinutto Nico N Bakshi Rohit R Bischof Antje A Calabresi Peter A PA Caverzasi Eduardo E Constable R Todd RT Datta Esha E Kirkish Gina G Nair Govind G Oh Jiwon J Pelletier Daniel D Pham Dzung L DL Reich Daniel S DS Rooney William W Roy Snehashis S Schwartz Daniel D Shinohara Russell T RT Sicotte Nancy L NL Stern William A WA Tagge Ian I Tauhid Shahamat S Tummala Subhash S Henry Roland G RG
Magnetic resonance in medicine 20170615 3
<h4>Purpose</h4>To explore (i) the variability of upper cervical cord area (UCCA) measurements from volumetric brain 3D T<sub>1</sub> -weighted scans related to gradient nonlinearity (GNL) and subject positioning; (ii) the effect of vendor-implemented GNL corrections; and (iii) easily applicable methods that can be used to retrospectively correct data.<h4>Methods</h4>A multiple sclerosis patient was scanned at seven sites using 3T MRI scanners with the same 3D T<sub>1</sub> -weighted protocol wi ...[more]