High versus low attenuation thresholds to determine the solid component of ground-glass opacity nodules.


ABSTRACT:

Objectives

To evaluate and compare the diagnostic accuracy of high versus low attenuation thresholds for determining the solid component of ground-glass opacity nodules (GGNs) for the differential diagnosis of adenocarcinoma in situ (AIS) from minimally invasive adenocarcinoma (MIA) and invasive adenocarcinoma (IA).

Methods

Eighty-six pathologically confirmed GGNs < 3 cm observed in 86 patients (27 male, 59 female; mean age, 59.3 ± 11.0 years) between January 2013 and December 2015 were retrospectively included. The solid component of each GGN was defined using two different attenuation thresholds: high (-160 Hounsfield units [HU]) and low (-400 HU). According to the presence or absence of solid portions, each GGN was categorized as a pure GGN or part-solid GGN. Solid components were regarded as indicators of invasive foci, suggesting MIA or IA.

Results

Among the 86 GGNs, there were 57 cases of IA, 19 of MIA, and 10 of AIS. Using the high attenuation threshold, 44 were categorized as pure GGNs and 42 as part-solid GGNs. Using the low attenuation threshold, 13 were categorized as pure GGNs and 73 as part-solid GGNs. The sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy for the invasive focus were 55.2%, 100%, 100%, 22.7%, and 60.4%, respectively, for the high attenuation threshold, and 93.4%, 80%, 97.2%, 61.5%, and 91.8%, respectively, for the low attenuation threshold.

Conclusion

The low attenuation threshold was better than the conventional high attenuation threshold for determining the solid components of GGNs, which indicate invasive foci.

SUBMITTER: Lee JH 

PROVIDER: S-EPMC6193644 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

High versus low attenuation thresholds to determine the solid component of ground-glass opacity nodules.

Lee Jae Ho JH   Kim Tae Hoon TH   Lee Sungsoo S   Han Kyunghwa K   Byun Min Kwang MK   Chang Yoon Soo YS   Kim Hyung Jung HJ   Lee Geun Dong GD   Park Chul Hwan CH  

PloS one 20181018 10


<h4>Objectives</h4>To evaluate and compare the diagnostic accuracy of high versus low attenuation thresholds for determining the solid component of ground-glass opacity nodules (GGNs) for the differential diagnosis of adenocarcinoma in situ (AIS) from minimally invasive adenocarcinoma (MIA) and invasive adenocarcinoma (IA).<h4>Methods</h4>Eighty-six pathologically confirmed GGNs < 3 cm observed in 86 patients (27 male, 59 female; mean age, 59.3 ± 11.0 years) between January 2013 and December 201  ...[more]

Similar Datasets

| S-EPMC7141214 | biostudies-literature
| S-EPMC10713293 | biostudies-literature
| S-EPMC7330407 | biostudies-literature
| S-EPMC9510793 | biostudies-literature
| S-EPMC8415279 | biostudies-literature
| S-EPMC8339807 | biostudies-literature
| S-EPMC10823549 | biostudies-literature
| S-EPMC7336435 | biostudies-literature
| S-EPMC7086215 | biostudies-literature