Unknown

Dataset Information

0

Evaluation of valve function in antireflux biliary metal stents.


ABSTRACT:

Background

To overcome duodenobiliary reflux induced by biliary stents, antireflux valve (ARV) biliary stents have been developed and showed improvement in stent patency. However, negative study results have also been reported because stent patency may be decreased by the malfunction of ARV itself. Given such mixed results, the true efficacy of ARV remains unknown and the mechanism of its dysfunction needs to be clearly elucidated. The aim of this study was to investigate the exact mechanism of ARV dysfunction using in vitro phantom models.

Methods

Two experimental models were designed to evaluate two important environmental factors suspected to cause ARV malfunction, i.e. bile flow and pH. Three types of ARV metal stents from different companies were used for the experiments: a funnel type ARV, a windsock type ARV, and a wine glass-shaped ARV. Ten stents of each type were tested (five stents in the bile flow phantom model, and another five stents in the duodenal pH environmental model). To determine ARV malfunction, ARV-induced flow resistance was measured using a custom-made testing device. All stents from the two models were removed every 2 weeks for 12 weeks after stent insertion and were evaluated on morphological and functional changes of the ARV.

Results

Only ARV of wine glass-shaped ARV was morphologically changed due to silicone bond detachment in the bile flow model. All types of ARV were morphologically changed in the pH model. The morphological changes of ARV influenced the flow resistance. The antegrade pressure gradients were increased over time in the pH model (p?ConclusionsMorphological change of the ARVs may induce dysfunction of ARV metal stents, which is mainly due to duodenal pH environment. In the future, development of new ARV that is not affected by duodenal environmental factors can be expected to improve stent patency.

SUBMITTER: Kwon CI 

PROVIDER: S-EPMC6194677 | biostudies-literature | 2018 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of valve function in antireflux biliary metal stents.

Kwon Chang-Il CI   Moon Jong Pil JP   Yun Ho H   Jeong Seok S   Koh Dong Hee DH   Lee Woo Jung WJ   Ko Kwang Hyun KH   Kang Dae Hwan DH  

BMC gastroenterology 20181019 1


<h4>Background</h4>To overcome duodenobiliary reflux induced by biliary stents, antireflux valve (ARV) biliary stents have been developed and showed improvement in stent patency. However, negative study results have also been reported because stent patency may be decreased by the malfunction of ARV itself. Given such mixed results, the true efficacy of ARV remains unknown and the mechanism of its dysfunction needs to be clearly elucidated. The aim of this study was to investigate the exact mecha  ...[more]

Similar Datasets

| S-EPMC7220473 | biostudies-literature
| S-EPMC5831319 | biostudies-literature
| S-EPMC3116134 | biostudies-literature
| S-EPMC7483763 | biostudies-literature
| S-EPMC8593498 | biostudies-literature
| S-EPMC7649830 | biostudies-literature
| S-EPMC4747571 | biostudies-literature
| S-EPMC8357580 | biostudies-literature
| S-EPMC4924429 | biostudies-literature
| S-EPMC5234016 | biostudies-literature