Unknown

Dataset Information

0

Comparison of the Vitek MS and Bruker Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Systems for Identification of Chryseobacterium Isolates from Clinical Specimens and Report of Uncommon Chryseobacterium Infections in Humans.


ABSTRACT: Matrix-assisted laser desorption ionization-time of flight mass spectrometry is becoming more popular and is replacing traditional identification methods in the clinical microbiology laboratory. We aimed to compare the Vitek mass spectrometry (MS) and Bruker Biotyper systems for the identification of Chryseobacterium isolated from clinical specimens and to report uncommon Chryseobacterium infections in humans. The microbial database from a hospital was searched for records between 2005 and 2016 to identify cultures that yielded Chryseobacterium Species identification by the Vitek MS and Bruker Biotyper systems was compared to identification by 16S rRNA gene sequencing. Over the study period, 140 Chryseobacterium isolates were included. Based on 16S rRNA gene sequencing, 78 isolates were C. indologenes, 39 were C. gleum, 12 were uncommon Chryseobacterium species (C. arthrosphaerae, C. culicis, C. cucumeris, C. bernardetii, C. artocarpi, and C. daecheongense), and the remaining 11 isolates were only identified at the genus level. The Vitek MS and Bruker Biotyper systems correctly identified 98.7% and 100% of C. indologenes isolates, respectively. While the Bruker Biotyper accurately identified 100% of C. gleum isolates, the Vitek MS system correctly identified only 2.6% of isolates from this species. None of the uncommon Chryseobacterium species were successfully identified by either of these two systems. The overall accuracies of Chryseobacterium identification at the species level by the Vitek MS and Bruker Biotyper systems were 60.5% and 90.7%, respectively. An upgrade and correction of the Vitek MS and Bruker Biotyper databases is recommended to correctly identify Chryseobacterium species.

SUBMITTER: Lin JN 

PROVIDER: S-EPMC6204688 | biostudies-literature | 2018 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of the Vitek MS and Bruker Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Systems for Identification of Chryseobacterium Isolates from Clinical Specimens and Report of Uncommon Chryseobacterium Infections in Humans.

Lin Jiun-Nong JN   Teng Shih-Hua SH   Lai Chung-Hsu CH   Yang Chih-Hui CH   Huang Yi-Han YH   Lin Hsiu-Fang HF   Lin Hsi-Hsun HH  

Journal of clinical microbiology 20181025 11


Matrix-assisted laser desorption ionization-time of flight mass spectrometry is becoming more popular and is replacing traditional identification methods in the clinical microbiology laboratory. We aimed to compare the Vitek mass spectrometry (MS) and Bruker Biotyper systems for the identification of <i>Chryseobacterium</i> isolated from clinical specimens and to report uncommon <i>Chryseobacterium</i> infections in humans. The microbial database from a hospital was searched for records between  ...[more]

Similar Datasets

| S-EPMC3697712 | biostudies-literature
| S-EPMC3318529 | biostudies-literature
| S-EPMC4844720 | biostudies-literature
| S-EPMC3716079 | biostudies-literature
| S-EPMC4400743 | biostudies-literature
| S-EPMC4097692 | biostudies-literature
| S-EPMC2238100 | biostudies-literature
| S-EPMC4809926 | biostudies-literature
| S-EPMC6224550 | biostudies-literature
| S-EPMC4767983 | biostudies-literature