Diagnostic performance of Elecsys immunoassays for cerebrospinal fluid Alzheimer's disease biomarkers in a nonacademic, multicenter memory clinic cohort: The ABIDE project.


ABSTRACT:

Introduction

We compared the automated Elecsys and manual Innotest immunoassays for cerebrospinal fluid (CSF) Alzheimer's disease biomarkers in a multicenter diagnostic setting.

Methods

We collected CSF samples from 137 participants in eight local memory clinics. Amyloid ?(1-42) (A?42), total tau (t-tau), and phosphorylated tau (p-tau) were centrally analyzed with Innotest and Elecsys assays. Concordances between methods were assessed.

Results

Biomarker results strongly correlated between assays with Spearman's ? 0.94 for A?42, 0.98 for t-tau, and 0.98 for p-tau. Using Gaussian mixture modeling, cohort-specific cut-points were estimated at 1092 pg/mL for A?42, 235 pg/mL for t-tau, and 24 pg/mL for p-tau. We found an excellent concordance of biomarker abnormality between assays of 97% for A?42 and 96% for both t-tau and p-tau.

Discussion

The high concordances between Elecsys and Innotest in this nonacademic, multicenter cohort support the use of Elecsys for CSF Alzheimer's disease diagnostics and allow conversion of results between methods.

SUBMITTER: Willemse EAJ 

PROVIDER: S-EPMC6215060 | biostudies-literature | 2018

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic performance of Elecsys immunoassays for cerebrospinal fluid Alzheimer's disease biomarkers in a nonacademic, multicenter memory clinic cohort: The ABIDE project.

Willemse Eline A J EAJ   van Maurik Ingrid S IS   Tijms Betty M BM   Bouwman Femke H FH   Franke Andreas A   Hubeek Isabelle I   Boelaarts Leo L   Claus Jules J JJ   Korf Esther S C ESC   van Marum Rob J RJ   Roks Gerwin G   Schoonenboom Niki N   Verwey Nicolaas N   Zwan Marissa D MD   Wahl Simone S   van der Flier Wiesje M WM   Teunissen Charlotte E CE  

Alzheimer's & dementia (Amsterdam, Netherlands) 20180912


<h4>Introduction</h4>We compared the automated Elecsys and manual Innotest immunoassays for cerebrospinal fluid (CSF) Alzheimer's disease biomarkers in a multicenter diagnostic setting.<h4>Methods</h4>We collected CSF samples from 137 participants in eight local memory clinics. Amyloid β(1-42) (Aβ42), total tau (t-tau), and phosphorylated tau (p-tau) were centrally analyzed with Innotest and Elecsys assays. Concordances between methods were assessed.<h4>Results</h4>Biomarker results strongly cor  ...[more]

Similar Datasets

| S-EPMC4995359 | biostudies-literature
| S-EPMC7505008 | biostudies-literature
| S-EPMC7110644 | biostudies-literature
| S-EPMC7592690 | biostudies-literature
| S-EPMC8122152 | biostudies-literature
| PRJNA856347 | ENA
| S-EPMC6667786 | biostudies-literature
| S-EPMC5318541 | biostudies-literature
| S-EPMC6974141 | biostudies-literature
| S-EPMC6466790 | biostudies-literature