Unknown

Dataset Information

0

Depth-Based, Motion-Stabilized Colorization of Microscope-Integrated Optical Coherence Tomography Volumes for Microscope-Independent Microsurgery.


ABSTRACT:

Purpose

We develop and assess the impact of depth-based, motion-stabilized colorization (color) of microscope-integrated optical coherence tomography (MIOCT) volumes on microsurgical performance and ability to interpret surgical volumes.

Methods

Color was applied in real-time as gradients indicating axial position and stabilized based on calculated center of mass. In a test comparing colorization versus grayscale visualizations of prerecorded intraoperative volumes from human surgery, ophthalmologists (N = 7) were asked to identify retinal membranes, the presence of an instrument, its contact with tissue, and associated deformation of the retina. In a separate controlled trial, trainees (N = 15) performed microsurgical skills without conventional optical visualization and compared colorized versus grayscale MIOCT visualization on a stereoptic screen. Skills included thickness identification, instrument placement, and object manipulation, and were assessed based on time, performance metrics, and confidence.

Results

In intraoperative volume testing, colorization improved ability to differentiate membrane from retina (P < 0.01), correctly identify instrument contact with membrane (P = 0.03), and retinal deformation (P = 0.01). In model microsurgical skills testing, trainees working with colorized volumes were faster (P < 0.01) and more correct (P < 0.01) in assessments of thickness for recessed and elevated objects, were less likely to inadvertently contact a surface when approaching with an instrument (P < 0.01), and uniformly more confident (P < 0.01 for each) in conducting each skill.

Conclusions

Depth-based colorization enables effective identification of retinal membranes and tissue deformation. In microsurgical skill testing, it improves user efficiency, and confidence in microscope-independent, OCT-guided model surgical maneuvers.

Translational relevance

Novel depth-based colorization and stabilization technology improves the use of intraoperative MIOCT.

SUBMITTER: Bleicher ID 

PROVIDER: S-EPMC6218157 | biostudies-literature | 2018 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Depth-Based, Motion-Stabilized Colorization of Microscope-Integrated Optical Coherence Tomography Volumes for Microscope-Independent Microsurgery.

Bleicher Isaac D ID   Jackson-Atogi Moseph M   Viehland Christian C   Gabr Hesham H   Gabr Hesham H   Izatt Joseph A JA   Toth Cynthia A CA  

Translational vision science & technology 20181101 6


<h4>Purpose</h4>We develop and assess the impact of depth-based, motion-stabilized colorization (color) of microscope-integrated optical coherence tomography (MIOCT) volumes on microsurgical performance and ability to interpret surgical volumes.<h4>Methods</h4>Color was applied in real-time as gradients indicating axial position and stabilized based on calculated center of mass. In a test comparing colorization versus grayscale visualizations of prerecorded intraoperative volumes from human surg  ...[more]

Similar Datasets

| S-EPMC5963505 | biostudies-literature
| S-EPMC7414702 | biostudies-literature
| S-EPMC5369296 | biostudies-literature
| S-EPMC8120705 | biostudies-literature
| S-EPMC4871076 | biostudies-other
| S-EPMC6991137 | biostudies-literature
| S-EPMC3537843 | biostudies-literature
| S-EPMC4497534 | biostudies-literature
| S-EPMC6039290 | biostudies-literature