ABSTRACT: Background
To investigate the effect of idiopathic epiretinal membrane (ERM) on the retinal-image quality and psychophysical contrast sensitivity function (CSF).Methods
Forty-four subjects with diagnosis of idiopathic unilateral ERM were enrolled in this cross-sectional observational clinical study. The fellow unaffected eyes were set as the control group. For retinal-image quality assessment, an Optical Quality Analysis System (OQAS) based on double-pass technique was used to evaluate objective scatter index (OSI) and Strehl ratio. For visual performance, the CSF under photopic condition was measured.Results
For retinal-image quality, the result of double-pass device revealed a significant lower Strehl ratio and larger OSI in the ERM eyes compared to the fellow eyes (all P?ConclusionEyes affected with ERM showed poorer retinal-image quality and visual performance than the normal eyes. Retinal-image quality measured by OQAS based on double-pass technique could be useful for assessing the retinal-image quality for ERM-affected eyes, in which retinal scattering was significantly increased.
SUBMITTER: Liu L
PROVIDER: S-EPMC6223062 | biostudies-literature | 2018 Nov
REPOSITORIES: biostudies-literature
Liu Limei L Wang Yi Y Liu Ju J Liu Wu W
BMC ophthalmology 20181107 1
<h4>Background</h4>To investigate the effect of idiopathic epiretinal membrane (ERM) on the retinal-image quality and psychophysical contrast sensitivity function (CSF).<h4>Methods</h4>Forty-four subjects with diagnosis of idiopathic unilateral ERM were enrolled in this cross-sectional observational clinical study. The fellow unaffected eyes were set as the control group. For retinal-image quality assessment, an Optical Quality Analysis System (OQAS) based on double-pass technique was used to ev ...[more]