Accelerated, free-breathing, noncontrast, electrocardiograph-triggered, thoracic MR angiography with stack-of-stars k-space sampling and GRASP reconstruction.


ABSTRACT: PURPOSE:To develop an accelerated, free-breathing, noncontrast, electrocardiograph-triggered, thoracic MR angiography (NC-MRA) pulse sequence capable of achieving high spatial resolution at clinically acceptable scan time and test whether it produces clinically acceptable image quality in patients with suspected aortic disease. METHODS:We modified a "coronary" MRA pulse sequence to use a stack-of-stars k-space sampling pattern and combined it with golden-angle radial sparse parallel (GRASP reconstruction to enable self-navigation of respiratory motion and high data acceleration. The performance of the proposed NC-MRA was evaluated in 13 patients, where clinical standard contrast-enhanced MRA (CE-MRA) was used as control. For visual analysis, two readers graded the conspicuity of vessel lumen, artifacts, and noise level on a 5-point scale (overall score index = sum of three scores). The aortic diameters were measured at seven standardized locations. The mean visual scores, inter-observer variability, and vessel diameters were compared using appropriate statistical tests. RESULTS:The overall mean visual score index (12.1 ± 1.7 for CE-MRA versus 12.1 ± 1.0 for NC-MRA) scores were not significantly different (P > 0.16). The two readers' scores were significantly different for CE-MRA (P = 0.01) but not for NC-MRA (P = 0.21). The mean vessel diameters were not significantly different, except at the proximal aortic arch (P < 0.03). The mean diameters were strongly correlated (R2 ? 0.96) and in good agreement (absolute mean difference ? 0.01 cm and 95% confidence interval ? 0.62 cm). CONCLUSION:This study shows that the proposed NC-MRA produces clinically acceptable image quality in patients at high spatial resolution (1.5 mm × 1.5 mm × 1.5 mm) and clinically acceptable scan time (~6 min).

SUBMITTER: Haji-Valizadeh H 

PROVIDER: S-EPMC6258265 | biostudies-literature | 2019 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accelerated, free-breathing, noncontrast, electrocardiograph-triggered, thoracic MR angiography with stack-of-stars k-space sampling and GRASP reconstruction.

Haji-Valizadeh Hassan H   Collins Jeremy D JD   Aouad Pascale J PJ   Serhal Ali M AM   Lindley Marc D MD   Pang Jianing J   Naresh Nivedita K NK   Carr James C JC   Kim Daniel D  

Magnetic resonance in medicine 20180905 1


<h4>Purpose</h4>To develop an accelerated, free-breathing, noncontrast, electrocardiograph-triggered, thoracic MR angiography (NC-MRA) pulse sequence capable of achieving high spatial resolution at clinically acceptable scan time and test whether it produces clinically acceptable image quality in patients with suspected aortic disease.<h4>Methods</h4>We modified a "coronary" MRA pulse sequence to use a stack-of-stars k-space sampling pattern and combined it with golden-angle radial sparse parall  ...[more]

Similar Datasets

| S-EPMC5675831 | biostudies-literature
| S-EPMC7722209 | biostudies-literature
| S-EPMC9892107 | biostudies-literature
| S-EPMC7718305 | biostudies-literature
| S-EPMC6031129 | biostudies-literature
| S-EPMC5814733 | biostudies-literature
| S-EPMC3198221 | biostudies-literature
| S-EPMC5545174 | biostudies-literature
| S-EPMC3818335 | biostudies-literature
| S-EPMC7479001 | biostudies-literature