Ontology highlight
ABSTRACT: Background
Congenital diaphragmatic hernia (CDH) is a fetal defect comprising an incomplete diaphragm and the herniation of abdominal organs into the chest cavity that interfere with fetal pulmonary development. Though the most promising treatment for CDH is via interventional fetoscopic tracheal occlusion (TO) surgery in-utero, it has produced mixed results due to the static nature of the inserted occlusion. We hypothesize that a suitable noninvasively-actuatable, cyclic-release tracheal occlusion device can be developed to enable dynamic tracheal occlusion (dTO) implementation.Objective
To conduct an in-vitro proof-of-concept investigation of the construction of thermo-responsive polymer valves designed for targeted activation within a physiologically realizable temperature range as a first step towards potential development of a noninvasively-actuatable implantable device to facilitate dynamic tracheal occlusion (dTO) therapy.Methods
Six thermo-responsive polymer valves, with a critical solution temperature slightly higher than normal physiological body temperature of 37°C, were fabricated using a copolymer of n-isopropylacrylamide (NIPAM) and dimethylacrylamide (DMAA). Three of the valves underwent ethylene oxide (EtO) sterilization while the other three served as controls for EtO-processing compatibility testing. Thermal response actuation of the valves and their steady-state flow performances were evaluated using water and caprine amniotic fluid.Results
All six valves consisting of 0.3-mole fraction of DMAA were tested for thermal actuation of caprine amniotic fluid flow at temperatures ranging from 30-44°C. They all exhibited initiation of valve actuation opening at ~40°C with full completion at ~44°C. The overall average coefficient of variation (CV) for the day-to-day flow performance of the valves tested was less than 12%. Based on a Student t-test, there was no significant difference in the operational characteristics for the EtO processed versus the non-EtO processed valves tested.Conclusions
We successfully fabricated and demonstrated physiological realizable temperature range operation of thermo-responsive polymer valves in-vitro and their suitability for standard EtO sterilization processing, a prerequisite for future in-vivo surgical implantation testing.
SUBMITTER: Baba JS
PROVIDER: S-EPMC6307748 | biostudies-literature | 2018
REPOSITORIES: biostudies-literature
PloS one 20181227 12
<h4>Background</h4>Congenital diaphragmatic hernia (CDH) is a fetal defect comprising an incomplete diaphragm and the herniation of abdominal organs into the chest cavity that interfere with fetal pulmonary development. Though the most promising treatment for CDH is via interventional fetoscopic tracheal occlusion (TO) surgery in-utero, it has produced mixed results due to the static nature of the inserted occlusion. We hypothesize that a suitable noninvasively-actuatable, cyclic-release trachea ...[more]