Unknown

Dataset Information

0

Acceptability, Accuracy, and Safety of Disposable Transnasal Capsule Endoscopy for Barrett's Esophagus Screening.


ABSTRACT: BACKGROUND & AIMS:Screening for Barrett's esophagus (BE) with conventional esophagogastroduodenoscopy (C-EGD) is expensive. We assessed the performance of a clinic-based, single use transnasal capsule endoscope (EG Scan II) for the detection of BE, compared to C-EGD as the reference standard. METHODS:We performed a prospective multicenter cohort study of patients with and without BE recruited from 3 referral centers (1 in the United States and 2 in the United Kingdom). Of 200 consenting participants, 178 (89%) completed both procedures (11% failed EG Scan due to the inability to intubate the nasopharynx). The mean age of participants was 57.9 years and 67% were male. The prevalence of BE was 53%. All subjects underwent the 2 procedures on the same day, performed by blinded endoscopists. Patients completed preference and validated tolerability (10-point visual analogue scale [VAS]) questionnaires within 14 days of the procedures. RESULTS:A higher proportion of patients preferred the EG Scan (54.2%) vs the C-EGD (16.7%) (P < .001) and the EG Scan had a higher VAS score (7.2) vs the C-EGD (6.4) (P = .0004). No serious adverse events occurred. The EG Scan identified any length BE with a sensitivity value of 0.90 (95% CI, 0.83-0.96) and a specificity value of 0.91 (95% CI, 0.82-0.96). The EG Scan identified long segment BE with a sensitivity value of 0.95 and short segment BE with a sensitivity values of 0.87. CONCLUSIONS:In a prospective study, we found the EG Scan to be safe and to detect BE with higher than 90% sensitivity and specificity. A higher proportion of patients preferred the EG Scan to C-EGD. This device might be used as a clinic-based tool to screen populations at risk for BE. ISRCTN registry identifier: 70595405; ClinicalTrials.gov no: NCT02066233.

SUBMITTER: Sami SS 

PROVIDER: S-EPMC6330075 | biostudies-literature | 2019 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Acceptability, Accuracy, and Safety of Disposable Transnasal Capsule Endoscopy for Barrett's Esophagus Screening.

Sami Sarmed S SS   Iyer Prasad G PG   Pophali Prachi P   Halland Magnus M   di Pietro Massimiliano M   Ortiz-Fernandez-Sordo Jacobo J   White Jonathan R JR   Johnson Michele M   Guha Indra Neil IN   Fitzgerald Rebecca C RC   Ragunath Krish K  

Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association 20180803 4


<h4>Background & aims</h4>Screening for Barrett's esophagus (BE) with conventional esophagogastroduodenoscopy (C-EGD) is expensive. We assessed the performance of a clinic-based, single use transnasal capsule endoscope (EG Scan II) for the detection of BE, compared to C-EGD as the reference standard.<h4>Methods</h4>We performed a prospective multicenter cohort study of patients with and without BE recruited from 3 referral centers (1 in the United States and 2 in the United Kingdom). Of 200 cons  ...[more]

Similar Datasets

| S-EPMC6924752 | biostudies-literature
| S-EPMC2906782 | biostudies-literature
| S-EPMC4154478 | biostudies-literature
| S-EPMC8576742 | biostudies-literature
| S-EPMC3052700 | biostudies-literature
| S-EPMC7611389 | biostudies-literature
| S-EPMC8715859 | biostudies-literature
| S-EPMC8748556 | biostudies-literature
| S-EPMC6047285 | biostudies-literature
| S-EPMC4441439 | biostudies-literature