Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device.


ABSTRACT: BACKGROUND:Respiratory rate is among the first vital signs to change in deteriorating patients. The aims of this study were to evaluate the accuracy of respiratory rate measurements using a specifically dedicated reflection-mode photoplethysmographic signal analysis in a pathological condition (PPG-RR) and to validate its implementation within medical devices. METHODS:This study is derived from a data mining project, including all consecutive patients admitted to our ICU (ReaSTOC study, ClinicalTrials.gov identifier: NCT02893462). During the evaluation phase of the algorithm, PPG-RR calculations were retrospectively performed on PPG waveforms extracted from the data warehouse and compared with RR reference values. During the prospective phase, PPG-RR calculations were automatically and continuously performed using a dedicated device (FreeO2, Oxynov, Québec, QC, Canada). In all phases, reference RR was measured continuously using electrical thoracic impedance and chronometric evaluation (Manual-RR) over a 30-s period. RESULTS:In total, 201 ICU patients' recordings (SAPS II 51.7?±?34.6) were analysed during the retrospective evaluation phase, most of them being admitted for a respiratory failure and requiring invasive mechanical ventilation. PPG-RR determination was available in 95.5% cases, similar to reference (22?±?4 vs. 22?±?5 c/min, respectively; p?=?1), and well correlated with reference values (R?=?0.952; p?

SUBMITTER: L'Her E 

PROVIDER: S-EPMC6340913 | biostudies-literature | 2019 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device.

L'Her Erwan E   N'Guyen Quang-Thang QT   Pateau Victoire V   Bodenes Laetitia L   Lellouche François F  

Annals of intensive care 20190121 1


<h4>Background</h4>Respiratory rate is among the first vital signs to change in deteriorating patients. The aims of this study were to evaluate the accuracy of respiratory rate measurements using a specifically dedicated reflection-mode photoplethysmographic signal analysis in a pathological condition (PPG-RR) and to validate its implementation within medical devices.<h4>Methods</h4>This study is derived from a data mining project, including all consecutive patients admitted to our ICU (ReaSTOC  ...[more]

Similar Datasets

| S-EPMC7146569 | biostudies-literature
| S-EPMC6363316 | biostudies-literature
| S-EPMC7442567 | biostudies-literature
| S-EPMC5548984 | biostudies-other
| S-EPMC9056464 | biostudies-literature
| S-EPMC6395464 | biostudies-literature
| S-EPMC4594125 | biostudies-literature
| S-EPMC6165317 | biostudies-literature
| S-EPMC10855057 | biostudies-literature
| S-EPMC10977037 | biostudies-literature