Accuracy and Operational Characteristics of Xpert Human Immunodeficiency Virus Point-of-Care Testing at Birth and Until Week 6 in Human Immunodeficiency Virus-exposed Neonates in Tanzania.


ABSTRACT: Background:Point-of-care (PoC) systems for early infant diagnosis (EID) may improve timely infant human immunodeficiency virus (HIV) management. Experiences within African public health settings are limited. Methods:We evaluated the accuracy and operational feasibility of the Xpert HIV-1 Qual for PoC-EID testing, using fresh blood and dried blood spots (DBS) samples at obstetric health facilities in Tanzania at birth and at postpartum weeks 1, 2, 3, and 6 in HIV-exposed infants. Test results were confirmed using TaqMan DBS HIV-deoxyribonucleic acid and/or plasma HIV-ribonucleic acid (RNA) testing. Results:At week 6, 15 (2.5%) out of 614 infants were diagnosed with HIV; 10 (66.7%) of them at birth (median HIV-RNA 4570 copies/mL). At birth, the Xpert-PoC and Xpert-DBS were 100% sensitive (95% confidence intervals: PoC, 69.2-100%; DBS, 66.4-100%) and 100% specific (PoC, 92.1-100%; DBS, 88.4-100%). By week 3, 5 infants with intra/postpartum HIV-infection (median HIV-RNA 1 160 000 copies/mL) were all correctly diagnosed by Xpert. In 2 cases, Xpert-PoC testing correctly identified HIV-infection when DBS tests (Xpert and TaqMan) were negative, suggesting a greater sensitivity. In 2 infants with confirmed HIV at birth, all tests were negative at week 6, possibly because of viral suppression under nevirapine prophylaxis. Problems were reported in 183/2736 (6.7%) of Xpert-PoC tests, mostly related to power cuts (57.9%). Conclusions:We demonstrated excellent Xpert HIV-1 Qual performance and good operational feasibility for PoC-EID testing at obstetric health facilities. Week 6 sensitivity issues were possibly related to nevirapine prophylaxis, supporting additional birth PoC-EID testing to avoid underdiagnosis. Clinical Trials Registration:NCT02545296.

SUBMITTER: Sabi I 

PROVIDER: S-EPMC6355822 | biostudies-literature | 2019 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy and Operational Characteristics of Xpert Human Immunodeficiency Virus Point-of-Care Testing at Birth and Until Week 6 in Human Immunodeficiency Virus-exposed Neonates in Tanzania.

Sabi Issa I   Mahiga Hellen H   Mgaya Jimson J   Geisenberger Otto O   Kastner Sabine S   Olomi Willyhelmina W   Saathoff Elmar E   Njovu Lilian L   Lueer Cornelia C   France John J   Maboko Leonard L   Ntinginya Nyanda Elias NE   Hoelscher Michael M   Kroidl Arne A  

Clinical infectious diseases : an official publication of the Infectious Diseases Society of America 20190201 4


<h4>Background</h4>Point-of-care (PoC) systems for early infant diagnosis (EID) may improve timely infant human immunodeficiency virus (HIV) management. Experiences within African public health settings are limited.<h4>Methods</h4>We evaluated the accuracy and operational feasibility of the Xpert HIV-1 Qual for PoC-EID testing, using fresh blood and dried blood spots (DBS) samples at obstetric health facilities in Tanzania at birth and at postpartum weeks 1, 2, 3, and 6 in HIV-exposed infants. T  ...[more]

Similar Datasets

| S-EPMC3813463 | biostudies-literature
| S-EPMC8563165 | biostudies-literature
| S-EPMC5407375 | biostudies-literature
| S-EPMC5183465 | biostudies-literature
| S-EPMC8589403 | biostudies-literature
| S-EPMC5564013 | biostudies-literature
| S-EPMC3173518 | biostudies-literature
| PRJNA387132 | ENA
| PRJNA279269 | ENA
| PRJNA735122 | ENA