Unknown

Dataset Information

0

Simulated Performance of the Cleveland Clinic Continuous-Flow Total Artificial Heart Using the Virtual Mock Loop.


ABSTRACT: Our new Virtual Mock Loop (VML) is a mathematical model designed to simulate the human cardiovascular system and gauge performance of mechanical circulatory support devices. We aimed to mimic the hemodynamic performance of Cleveland Clinic's self-regulating continuous-flow total artificial heart (CFTAH) via VML and evaluate VML's accuracy versus bench data from our standard mock circulatory loop. The VML reproduced 23 hemodynamic conditions. Systemic/pulmonary vascular resistances and pump rotational speed were set for VML from bench test data. We compared outputs (pump flow, left/right pump pressure rise, normalized pump performance, and atrial pressure difference) of the two methods. Data from pump flow and left pump pressure rise were similar, but right pump pressure rise slightly differed. Left pump normalized pump performance curves were similar. Right pump VML results were within the same performance range indicated by bench tests. The plots of atrial pressure differences of VML versus bench-test data were similar, but slightly differed in the midrange of systemic/pulmonary gradients. Virtual Mock Loop successfully reproduced results from our mock circulatory loop of CFTAH test conditions. The CFTAH's self-regulation feature of right pump performance was also calculated effectively. We foresee using versions of the VML for training, simulating physiologic cardiac conditions, and patient monitoring.

SUBMITTER: Miyamoto T 

PROVIDER: S-EPMC6359994 | biostudies-literature | 2019 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Simulated Performance of the Cleveland Clinic Continuous-Flow Total Artificial Heart Using the Virtual Mock Loop.

Miyamoto Takuma T   Horvath David J DJ   Horvath Dennis W DW   Karimov Jamshid H JH   Byram Nicole N   Kuban Barry D BD   Fukamachi Kiyotaka K  

ASAIO journal (American Society for Artificial Internal Organs : 1992) 20190801 6


Our new Virtual Mock Loop (VML) is a mathematical model designed to simulate the human cardiovascular system and gauge performance of mechanical circulatory support devices. We aimed to mimic the hemodynamic performance of Cleveland Clinic's self-regulating continuous-flow total artificial heart (CFTAH) via VML and evaluate VML's accuracy versus bench data from our standard mock circulatory loop. The VML reproduced 23 hemodynamic conditions. Systemic/pulmonary vascular resistances and pump rotat  ...[more]

Similar Datasets

| S-EPMC6647019 | biostudies-literature
| S-EPMC9874141 | biostudies-literature
| S-EPMC7164989 | biostudies-literature
| S-EPMC9802556 | biostudies-literature
| S-EPMC4414704 | biostudies-literature
| S-EPMC4554829 | biostudies-literature
| S-EPMC8178460 | biostudies-literature
| S-EPMC4626017 | biostudies-literature
| S-EPMC8343392 | biostudies-literature
2015-11-30 | E-MTAB-3131 | biostudies-arrayexpress