Comparison of displacement of polyvinyl chloride and silicone leftsided double-lumen tubes during lateral positioning.


ABSTRACT: BACKGROUND:Compared to an equivalent sized polyvinyl chloride (PVC) double-lumen tube (DLT), a silicone DLT has a shorter endobronchial segment. The aim of this study was to compare the incidence of clinically significant displacement of left-sided PVC and silicone DLTs after a positional change to a lateral position from a supine position and determine its effect on the need for DLT repositioning for successful lung separation in patients undergoing thoracic surgery. METHODS:One hundred eight adult patients requiring one-lung ventilation were randomly divided into group P (PVC DLT, n = 54) and group S (Silicone DLT, n = 54). The position of the DLT was observed before and after lateral positioning to assess the effect of the position change. The incidence of clinically significant displacement (> 10 mm) of DLT was compared between the groups. RESULTS:DLTs were clinically significantly displaced in group P (17/48, 35.4%) and group S (18/52, 34.6%) after lateral positioning (P = 0.933). A proximal displacement (31.3% [group P] and 25.0% [group S]) was more common than distal displacement (4.2% [group P] and 9.6% [group S]), with no significant intergroup difference (P = 0.494). After lateral positioning, critical malposition of DLT with bronchial herniation to the right main bronchus was 8.3% (group P) and 7.9% (group S, P = 0.906). CONCLUSIONS:Left-sided PVC and silicone DLTs produced comparable incidences of clinically significant displacement and, consequently, required similar rates of repositioning for successful lung separation after lateral positioning.

SUBMITTER: Kwon NK 

PROVIDER: S-EPMC6369336 | biostudies-literature | 2019 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of displacement of polyvinyl chloride and silicone leftsided double-lumen tubes during lateral positioning.

Kwon Nyeong Keon NK   Jung Sung Mee SM   Park Sang-Jin SJ   Han Yonghee Y  

Korean journal of anesthesiology 20180620 1


<h4>Background</h4>Compared to an equivalent sized polyvinyl chloride (PVC) double-lumen tube (DLT), a silicone DLT has a shorter endobronchial segment. The aim of this study was to compare the incidence of clinically significant displacement of left-sided PVC and silicone DLTs after a positional change to a lateral position from a supine position and determine its effect on the need for DLT repositioning for successful lung separation in patients undergoing thoracic surgery.<h4>Methods</h4>One  ...[more]

Similar Datasets

| S-EPMC10244602 | biostudies-literature
| S-EPMC7230200 | biostudies-literature
| S-EPMC7034193 | biostudies-literature
| S-EPMC8070095 | biostudies-literature
| S-EPMC9175383 | biostudies-literature
| S-EPMC6011323 | biostudies-literature
| S-EPMC9919178 | biostudies-literature
| S-EPMC4799618 | biostudies-literature
| S-EPMC9641341 | biostudies-literature
| S-EPMC6475141 | biostudies-literature