Comparison of Six Automated Immunoassays With Isotope-Diluted Liquid Chromatography-Tandem Mass Spectrometry for Total Thyroxine Measurement.


ABSTRACT: BACKGROUND:Accurate serum total thyroxine (TT4) measurement is important for thyroid disorder diagnosis and management. We compared the performance of six automated immunoassays with that of isotope-diluted liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) as the reference method. We also evaluated the correlation of thyroid stimulating hormone (TSH) with TT4 measured by ID-LC-MS/MS and immunoassays. METHODS:Serum was collected from 156 patients between October 2015 and January 2016. TT4 was measured by immunoassays from Abbott (Architect), Siemens (ADVIA Centaur XP), Roche (E601), Beckman-Coulter (Dxi800), Autobio (Autolumo A2000), and Mindray (CL-1000i), and by ID-LC-MS/MS. Results were analyzed using Passing-Bablok regression and Bland-Altman plots. Minimum requirements based on biological variation were as follows: a mean bias of ?4.5% and total imprecision (CV) of ?3.7%. RESULTS:All immunoassays showed a correlation >0.945 with ID-LC-MS/MS; however, the slope of the Passing-Bablok regression line varied from 0.886 (Mindray) to 1.23 (Siemens) and the intercept from -12.8 (Siemens) to 4.61 (Mindray). Only Autobio, Beckman-Coulter, and Roche included the value of one in the 95% confidence interval for slope. The mean bias ranged from -10.8% (Abbott) to 9.0% (Siemens), with the lowest value noted for Roche (3.5%) and the highest for Abbott (-10.8%). Only Abbott and Roche showed within-run and total CV ?3.7%. CONCLUSIONS:Though all immunoassays correlated strongly with ID-LC-MS/MS, most did not meet the minimum clinical requirement. Laboratories and immunoassay manufacturers must be aware of these limitations.

SUBMITTER: Yu S 

PROVIDER: S-EPMC6400722 | biostudies-literature | 2019 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of Six Automated Immunoassays With Isotope-Diluted Liquid Chromatography-Tandem Mass Spectrometry for Total Thyroxine Measurement.

Yu Songlin S   Zhou Weiyan W   Cheng Xinqi X   Meng Qinghui Q   Li Honglei H   Hou Li'an L   Lu Jun J   Xie Shaowei S   Cheng Qian Q   Zhang Chuanbao C   Qiu Ling L  

Annals of laboratory medicine 20190701 4


<h4>Background</h4>Accurate serum total thyroxine (TT4) measurement is important for thyroid disorder diagnosis and management. We compared the performance of six automated immunoassays with that of isotope-diluted liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) as the reference method. We also evaluated the correlation of thyroid stimulating hormone (TSH) with TT4 measured by ID-LC-MS/MS and immunoassays.<h4>Methods</h4>Serum was collected from 156 patients between October 2015 and  ...[more]

Similar Datasets

| S-EPMC4366052 | biostudies-literature
| S-EPMC4835639 | biostudies-literature
| S-EPMC8671195 | biostudies-literature
| S-EPMC7017602 | biostudies-literature
| S-EPMC7180471 | biostudies-literature
| S-EPMC2652408 | biostudies-literature
| S-EPMC7339657 | biostudies-literature
| S-EPMC4823009 | biostudies-literature
| S-EPMC9467846 | biostudies-literature