Short-acquisition-time JPRESS and its application to paediatric brain tumours.


ABSTRACT:

Objective

To develop and assess a short-duration JPRESS protocol for detection of overlapping metabolite biomarkers and its application to paediatric brain tumours at 3 Tesla.

Materials and methods

The short-duration protocol (6 min) was optimised and compared for spectral quality to a high-resolution (38 min) JPRESS protocol in a phantom and five healthy volunteers. The 6-min JPRESS was acquired from four paediatric brain tumours and compared with short-TE PRESS.

Results

Metabolite identification between the 6- and 38-min protocols was comparable in phantom and volunteer data. For metabolites with Cramer-Rao lower bounds > 50%, interpretation of JPRESS increased confidence in assignment of lactate, myo-Inositol and scyllo-Inositol. JPRESS also showed promise for the detection of glycine and taurine in paediatric brain tumours when compared to short-TE MRS.

Conclusion

A 6-min JPRESS protocol is well tolerated in paediatric brain tumour patients. Visual inspection of a 6-min JPRESS spectrum enables identification of a range of metabolite biomarkers of clinical interest.

SUBMITTER: Carlin D 

PROVIDER: S-EPMC6424926 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

2024-03-21 | GSE247880 | GEO
2013-07-29 | GSE42658 | GEO
2015-12-10 | GSE42657 | GEO
2013-07-29 | GSE42656 | GEO
| S-ECPF-GEOD-42658 | biostudies-other
2013-07-29 | E-GEOD-42658 | biostudies-arrayexpress
2013-07-29 | E-GEOD-42656 | biostudies-arrayexpress
2015-12-10 | E-GEOD-42657 | biostudies-arrayexpress
| S-EPMC6824674 | biostudies-literature
| S-EPMC6365981 | biostudies-literature