Unknown

Dataset Information

0

Wirelessly Controlled Implantable System for On-demand and Pulsatile Insulin Administration.


ABSTRACT: We propose a wirelessly controlled implantable system for on-demand and pulsatile insulin delivery with a more convenient and safer strategy than currently available strategies. The system is a combined entity of a magnetically driven pump (i.e., an MDP), external control device (i.e., an ECD) and mobile app. The MDP for implantation consists of a plunger, barrel and drug reservoir, where an accurate amount of insulin can be infused in a pulsatile manner only at the time when a magnetic force is applied to actuate the plunger in the barrel. The ECD at the outside body can modulate the MDP actuation with an electromagnet and its control circuit, and this modulation can be wirelessly controlled by the mobile app. As a safety feature, the mobile app is programmed to pre-set the restrictions for the insulin dose and administration schedule to avoid overdose. The system is shown to infuse insulin in a highly reproducible manner, but it does not allow for insulin infusion when the pre-set restrictions are violated. When tested with diabetic rats, the profiles of insulin plasma concentration and blood glucose level are similar to those of animals treated with a subcutaneous injection of the same dose of insulin.

SUBMITTER: Lee SH 

PROVIDER: S-EPMC6428930 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC5399301 | biostudies-literature
| S-EPMC4322607 | biostudies-literature
| S-EPMC7775752 | biostudies-literature
| S-EPMC4786798 | biostudies-other
| S-EPMC7764162 | biostudies-literature
| S-EPMC8626078 | biostudies-literature
| S-EPMC8336080 | biostudies-literature
| S-EPMC7202567 | biostudies-literature