Unknown

Dataset Information

0

The unstable pacing thresholds of the leadless transcatheter pacemaker affected by body positions in subacute phase after implant.


ABSTRACT:

Background

If the threshold at implant of leadless transcatheter pacemakers (LTPs) is less than 2.0?V, pacing thresholds reportedly decrease significantly by 1?month and maintain an optimal value of less than 1.0?V by 6?months.

Case summary

We report a case series of two patients with unstable pacing thresholds of the LTPs in the subacute phase after implant. The first patient (77-year-old man) was implanted an LTP for sick sinus syndrome. At that time of implant, the pacing threshold was 0.9?V at 0.24?ms. At 1?week and 1?month later, the threshold had increased to more than 2.0?V at 0.24?ms. We investigated the trend data for the week and found variations in the threshold. The second patient (81-year-old man) was implanted an LTP for bradycardia and atrial fibrillation. The pacing threshold at implantation was 0.63?V at 0.24?ms. One week later, the threshold had increased in supine position and decreased in sitting position. The trend data for the week were fluctuating greatly.

Discussion

The pacing threshold may increase to more than 2.0?V with significant fluctuation on assessment at 1?week and 1?month after implantation in association with changes in body position, even though we confirmed a stable threshold at implant. If an increased threshold is observed, it is necessary to check the trend data and threshold in each body position.

SUBMITTER: Yoh M 

PROVIDER: S-EPMC6439362 | biostudies-literature | 2019 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

The unstable pacing thresholds of the leadless transcatheter pacemaker affected by body positions in subacute phase after implant.

Yoh Masue M   Takagi Masahiko M   Takahashi Hiroki H   Yoshio Takuro T   Shiojima Ichiro I  

European heart journal. Case reports 20181227 1


<h4>Background</h4>If the threshold at implant of leadless transcatheter pacemakers (LTPs) is less than 2.0 V, pacing thresholds reportedly decrease significantly by 1 month and maintain an optimal value of less than 1.0 V by 6 months.<h4>Case summary</h4>We report a case series of two patients with unstable pacing thresholds of the LTPs in the subacute phase after implant. The first patient (77-year-old man) was implanted an LTP for sick sinus syndrome. At that time of implant, the pacing thres  ...[more]

Similar Datasets

| S-EPMC4589655 | biostudies-literature
| S-EPMC8360010 | biostudies-literature
| S-EPMC6850455 | biostudies-literature
| S-EPMC4876985 | biostudies-literature
| S-EPMC6009778 | biostudies-literature